Construction of a Silver(I) Ion-Selective Electrode Using Amine Phenol Ligand as Carrier and the Selective Determination of Silver in Actual Samples
Construction of a Silver(I) Ion-Selective Electrode Using Amine Phenol Ligand as Carrier and the Selective Determination of Silver in Actual Samples
Journal of the Korean Chemical Society. 2011. Feb, 55(1): 63-69
Copyright © 2011, The Korean Chemical Society
  • Received : October 27, 2010
  • Accepted : December 14, 2010
  • Published : February 20, 2011
Download
PDF
e-PUB
PubReader
PPT
Export by style
Article
Author
Metrics
Cited by
TagCloud
About the Authors
Wen-Ju Xu
Ya-Qin Chai
Ruo Yuan
yuanruo@swu.edu.cn

Abstract
This work discusses the fabrication, development and potential response behaviors of Ag + ion-selective electrodes (Ag + -ISE) based on N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane (L1) and N,N’-bis(2-hydroxybenzyl)-2,2-dimethyl-1,3-diaminopropane (L2) as carriers. The observations indicated that the resulting electrode based on L1 toward Ag + showed stable near-Nernst slope approaching 58.7 mV/dec and the optimum potential response characteristics in a linear range at least five orders of magnitude with a detection limit of 1.0×10 -6 M. The proposed electrode displayed the preferential selectivity to Ag + against other tested cations. The excellent potential analytical characteristics could lead to the successful applications of silver assay in significant real samples, indicating that the proposed Ag + -ISE showed a significant advancement of measurement capabilities. But for the electrode based on L2, the poor potential response characteristics were observed in total experiment process.
Keywords
INTRODUCTION
Potentiometric detectors based on ion-selective electrodes (ISEs) are especially suitable for fast, accurate, and selective detection of various target ions, because ionselective electrodes can offer high selectivity and sensitivity and are simple, handy and inexpensive as compared with other analytical instruments. Up to now, the design and synthesis of new carriers as the active components in ISEs have been received extensive attentions. 1 It was delightful that many effective compounds have been successfully used as suitable carriers in ISEs which were found to be sensitive to various alkali, alkaline earth and transition metal ions, such as Na + , K + , Mg 2+ , Ca 2+ , Ba 2+ , Hg 2+ , Cu 2+ , Ni 2+ , Pb 2+ and etc. 2 - 10
Correspondingly, the same interest has been focused on the fabrication and development for Ag + -ISEs. Some polymeric membrane and coated graphite Ag + -ISEs based on diffferent sensing materials as carriers have been successively reported in literatures. 11 - 20 However, most of the Ag + -ISEs seemed to be mainly based on crown ether derivatives and calixarene derivatives as carriers. 11 - 17 And those derivatives with functional groups containing nitrogen or sulfur atoms as donors were found to show particular selectivity toward soft heavy metals ions, such as Ag + , and Hg 2+ ion. 11 , 12 , 14 - 16 Thus, such Ag + -ISEs inevitably suffer serious interference from Hg 2+ ion. Moreover, complicated and tedious synthesis of crown ether derivatives and calixarene derivatives limited the analytical applications of these Ag + -ISEs.
As well known, the Schiff bases can easily form stable complexes with transition metal ions due to the performances of giving geometric and cavity control of hostguest complexation and modulation of lipophilicity, which can lead to their remakable sensitivity and selectivity for a specific ion. And almost all of metals form 1:1 metal complexes with them. Therefore, the applications and investigations for Schiff bases as sensing materials in ISEs have received increasing attentions. The Schiff bases with nitrogen and oxygen as donor atoms, which are well known to form strong complexes with transition metal ions, have been used as the carriers in ISEs which showed preferential selectivity to various metal ions, such as Ca 2+ , Cu 2+ , Ni 2+ , Mn 2+ , Pb 2+ , Fe 3+ , Cr 3+ and Cd 2+ . 6 , 8 , 10 , 21 - 25 However, Ag + -ISEs based on the Schiff bases as carriers have received little attentions. 17 In this work the easily synthesized and simple amine phenol (i.e. Schiff base) ligands, N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane (L1) and N, N’-bis(2-hydroxybenzyl)-2, 2-dimethyl-1, 3-diaminopropane (L2) with N and O as donor atoms, were used as the sensing components in ISEs. We found that the resulting electrode based on L1 as a simple and an effective carrier gave preferential selectivity and high sensitivity toward silver(I) ion against other tested cations. To our knowledge, this is the first time for the amine phenol ligands as the carriers in ISEs.
EXPERIMENTAL
- Reagents and apparatus
The amine phenol ligands used in this work were prepared according to the literature. 26 1,3-diaminopropane was obtained from Alfa Aesar China (Tianjin, China) Co. Ltd. 2,2-dimethyl-1,3-diaminopropane was purchased from Sigma-Aldrich. o -Nitrophenyl octyl ether ( o -NPOE) was synthesized according to the literature method. 27 Salicylaldehyde, poly vinyl chloride (PVC), sodium tetraphenylborate (NaTPB), ammonium and metal nitrates and other reagents were of analytical grade and available from Shanghai Chemical Reagent Co. (Shanghai, China). All aqueous solutions were prepared with deionized distilled water.
All potential measurements were performed by using a MP230 pH meter (Mettler Toledo, Switzerland) and a pHS-3C digital ion analyzer (Shanghai Dazhong Analytical Instruments, Shanghai, China). The conductance of solutions was measured with a conductivity meter (DDBJ-350, Shanghai Instrument Co., Shanghai, China). A conventional pH glass electrode was used to monitor pH values of working solutions. A double-junction-type Hg-Hg 2 Cl 2 electrode connected to the solution via a salt bridge containing 0.1 M KNO 3 solution was used as the external reference electrode.
- Electrodes fabrication
The general procedure for the ion-sensing polymeric membranes, according to Thomas, etc. 28 is as follows: amine phenol ligand (10.0 mg), plasticizer (200.0 mg), PVC (100.0 mg) and NaTPB (1.0 mg) were dissolved in 3 mL of freshly distilled THF. The resulting mixture was poured in a 3.5× 3.5 cm 2 glass plate, and then an elastic, semitransparent membrane with a thickness of approximate 0.2 mm was formed after gradual evaporation of the THF at room temperature (usually 24 h). The resulting membrane consisted of 3.22 wt% carrier, 64.31 wt% plasticizer, 32.15 wt% PVC and 0.32 wt% NaTPB. A smaller disc of about 8 mm diameter was cut from each master membrane and fixed into an electrode body. After that, 0.1 M KNO 3 +0.01 M AgNO 3 were used as the internal filling solution. Conditioning of the electrodes was done by soaking them in 0.01 M AgNO 3 .
- Potential measurements
The electrochemical cell was as following: Hg-Hg 2 Cl 2 |KCl(sat.)|0.1 M KNO 3 |sample solution|PVC-membrane| 0.1M KNO 3 +0.01M AgNO 3 |Ag-AgCl. The performance of the electrode was examined by measuring the cell emf for aqueous solutions of given ions. All emf values were measured at room temperature. To obtain the calibration curve of potential response, the data were constructed by plotting the potential ( E /mV) vs. the logarithm of the concentration of silver(I) ion species (log[Ag + ]) in tested solutions.
According to IUPAC, selectivity coefficients,
PPT Slide
Lager Image
were determined by the fixed interference method (FIM). 29 10 -2 M concentration of interfering cations was kept in various AgNO 3 solutions (pH 4.0) with concentrations of 10 -7 -10 -1 M.
RESULTS AND DISSCUSSION
- Preliminary study
In preliminary experiments, the complexation of carriers L1 and L2 with different cations was investigated conductometrically in dimethyl sulfoxide solution at 25 ℃, 30 in order to obtain a clue about the stability of the resulting complexes. The results obtained by the basic experiments showed addition of the ligand to the metal ion solution caused a continuous decrease in the molar conductance, and the decrease began to level off at molar ratios up to 1. The much sharper inflection point in the case of L1 relative to that for L2 was also observed. This indicated that the 1:1 complexes (metal ion:L1 or L2) may be formed in solutions and the complex of metal ion:L1 was more stable than metal ion:L2. Further, the formation constants (log K f ) of the resulting complex (Ag + :L1) obtained is 5.51±0.06, and it is 3.23±0.05 for the complex (Cu 2+ :L1). But for the other complexes formed by Pb 2+ , Co 2+ , Hg 2+ , Cr 3+ , Al 3+ , Mg 2+ , K + , Cd 2+ , Zn 2+ , Ba 2+ and L1, log K f is between 2.77 and 2.08. Therefore the most stable complex of L1 with Ag + reveals the much higher selectivity of L1 for Ag + over other cations, indicating that L1 is expected to act as a suitably selective carrier for constructing Ag + -ISE.
- Potential response characteristics of the electrodes
For carriers L1 and L2, some membrane electrodes containing anion additive NaTPB and plasticized with o -NPOE were prepared. The potential response calibration curves of the resulting electrodes toward Ag + are compared in . 1 . It can be seen that the electrode based on L1 displayed the optimum potential response to Ag + . This is likely due to the suitable size of the C-shaped cavity of the ligand L1, 25 which can lead to the effective formation of the affinities. The sub-Nernst sensitivity to Ag + for the L2-based electrode, on the other hand, might be attributed to the unfit size of L2 and larger steric hindrance for Ag + , which in turn cause the difficult complexing of L2 and Ag + and the inefficient cation exchange in the membrane interface.
Potential response characteristics of the electrodes based on carriers L1 and L2 toward Ag+(o-NPOE:PVC = 2:1 (w/w))
PPT Slide
Lager Image
Potential response characteristics of the electrodes based on carriers L1 and L2 toward Ag+ (o-NPOE:PVC = 2:1 (w/w))
PPT Slide
Lager Image
The calibration curves of the potential response of the electrodes based on L1 and L2 as carriers.
The more detailed analytical characteristics of the electrodes based on L1 and L2 are summarized in 1 . Clearly, 10.0 mg carrier in the membrane is the best helpful to achieve the optimum potential response for L1 and L2. The electrode B based on L1 displayed near-Nernst response to Ag + in the dynamic concentration range of 1.0×10 -1 -2.1×10 -6 M with the slope of 58.7±1.0 mV/dec and the IUPAC-defined detection limit of 1.0×10 -6 M. 29 The L2-based electrode E did not give good potential response, such as comparably low slope of the calibration curve (52.7±1.0 mV/dec) and narrow linear range about four orders of magnitude. Here, the Ag + -ISE based on the amine phenol L1 as simple and effective carrier showed better potential analytical characteristics, such as near-Nernst slope, wider linear range and lower detection limit, compared with the electrodes based on crown ether derivatives and calixarene derivatives. 11 , 12 , 14 - 16
IUPAC definition of response time is the time when 90% of the final value of the potential has been reached upon a change in concentration. 29 For the optimized electrode B, at Ag + concentrations above 1.0×10 -4 M, 60-80 s was sufficient to reach less than 0.3 mV/min change in the potential, while about 2-3 min was required for lower concentrations. When the electrode that had been exposed to 0.1 M AgNO 3 was immersed into a buffer solution without AgNO 3 , its potential returned to the baseline potential within 3 min. Thus, the reversibility of the fast potential response of the proposed electrode B to Ag + was suggested according to the above observations.
When the electrode B was immersed into 1.0×10 -2 M AgNO 3 solution over a period of 10 h, 0.67 mV ( n =40) of the standard deviation (SD) of the potential readings was observed. The potential readings for the same electrode dipped alternately into solutions of 1.0×10 -3 and 1.0×10 -2 M AgNO 3 gave a SD of 0.81. For the proposed electrode which was successively traced over 10 weeks, no detectable loss of the potential response characteristics was observed.
The effect of pH on the potential response was tested by immersing the electrode B into 1.0×10 -2 M AgNO 3 solution with different pH values (1.0-9.0). As shown in . 2 , pH range of 3.0-7.5 was suitable for the proposed electrode, and instable potential response behaviors were observed at lower (<3.0) or higher pH (>7.5) values.
PPT Slide
Lager Image
The potential response curves of the electrode B in 1.0×10–2 M AgNO3 solutions at different pH values.
- Selectivity
For the electrode B, the logarithm of selectivity coefficients,
PPT Slide
Lager Image
is summarized in . 3 . Obviously, the optimized electrode gave a high selectivity toward Ag + ion over other tested alkali and alkaline-earth metal ions, some transition metal ions and even a soft metal ion, Hg 2+ . Conventional Ag + -ISEs based on Ag 2 S generally suffer from terrible interference from Hg 2+ . Definitely, the present electrode based on the amine phenol ligand L1 is superior to the conventional Ag 2 S-based electrodes from this point of view.
PPT Slide
Lager Image
Selectivity coefficients, log KpotAg+,j, of the electrode B determined by FIM.
On the other hand, it should be noted that there was some interference of a bivalent metal ion, Cu 2+ , according to the selectivity coefficient
PPT Slide
Lager Image
indicating that Cu 2+ ion is the stronggest interferent. But the selectivity of an ion-selective electrode is concentration dependent. In order to establish the Cu 2+ concentration that can be tolerated in determining Ag + ion, the potential response of the electrode B was investigated in the 1.0×10 -7 -1.0 ×10 -1 M AgNO 3 solutions containing different concentrations of Cu 2+ ion. From . 4 , the presence of Cu 2+ at a concentration of 5.0×10 -5 M resulted in no deviation in the potential calibration curve ( E vs log[Ag + ]) compared with that obtained in pure Ag + solutions. Thus, the elelctrode B can tolerate Cu 2+ at concentration 5.0×10 -5 M over the whole dynamic concentration ranges. But Cu 2+ ion at higher concentrations caused a decrease of linear range and an increase of detection limit. So, Ag + can be determined in the concentration ranges larger than 8.3×10 -6 , 4.5×10 -5 , 8.9×10 -5 M in the presence of 1.0×10 -4 , 5.0×10 -4 and 1.0×10 -3 M Cu 2+ ion, respectively, which sugggested that the electrode B could be used to determine Ag + content in the presence of Cu 2+ ion of at least 10-fold concentration.
PPT Slide
Lager Image
Potential response curves of the electrode B toward Ag+ in the presence of different concentrations of Cu2+: 1.0×10–3, 5.0×10–4, 1.0×10–4, 5.0×10–5, and 1.0×10–5 M.
- Preliminary applications
The proposed electrode B was found to work well under laboratory conditions. It was used as an indicator to determine the end-point in the direct potential titration of 1.0×10 -3 M KCl solution with 5.0×10 -3 M AgNO 3 . As shown in . 5 , the resulting titration curve gave a relatively sharp inflection break at inflexion point.
PPT Slide
Lager Image
Potential titration curves of 1.0×10–3 M KCl solution with 5.0×10–3 M AgNO3 using electrode B as the indicator one.
In order to investigate the utility of the proposed Ag + -ISE in medicinal analysis, the potential response toward drug preparations was also investigated. Vitamin B 1 (V B1 ) is a chloride salt which has two chloride ions in a Vitamin B 1 molecule. The potential titration using the electrode B as the indicator one could be applied to the determination of the V B1 content in Vitamin B 1 tablets. The resulting titration curve is shown in . 6 . The content ( w ) of V B1 obtained was 15.22% ± 0.5 ( n =3). For comparison, the content of V B1 given by the conventional titration method, 31 in which the standard AgNO 3 solution as the titrant and K 2 CrO 4 as the indicator, was 15.17%±0.5 ( n =3).
The applicability of the electrode is further illustrated by measuring the Ag + ion potentiometrically in doubly distilled deionised water spiked with 40.0, 20.0 and 10.0 mg L -1 Ag + ion. The Ag + concentrations were determined by a standard curve method and atomic absorption spectrometric (AAS), respectively. The statistical comparison for the results obtained by the two methods was performed concerning precision and accuracy by using t test and F test. The results shown in 2 were found in good agreement. Moreover, at 95% confidence level ( f = 3), the obtained t value and F value don’t exceed the critical t value (3.18) and F value (9.28), indicating that there is no significant difference between the two methods with regard to precision and accuracy.
PPT Slide
Lager Image
Potential titration curves of Vitamin B1 solution with AgNO3 standard solution using electrode B as the indicator one.
Analysis of water samples spiked with silver(I) ion using the Ag+-ISE and AAS
PPT Slide
Lager Image
aAverage of four replicate measurements±standard deviation (SD).
Moreover, to demonstrate the analytical usefulness of the proposed Ag + -ISE in environmental analysis, it was used for silver content assay in waste fixing solution for photography (Samples 1 and 2 in 3 ) and electroplating wastewater (Samples 3-5 in 3 ) collected from local Xianghui Photographyshop and Electroplating Factory, respectively. The silver content was determined by using Gran’s plots method 32 with the electrode B as the indicator and compared with that given by AAS. The statistical comparison for the results was performed with regard to precision and accuracy by using t test and F test. From 3 , we can see that the silver content in waste fixing solution and electroplating wastewater given by the two methods shows good agreement. At 95% confidence level ( f = 3), the lower t value and F value than the critical t value (3.18) and F value (9.28) indicates that there is no significant difference between the two methods concerning precision and accuracy. The above observations display that the proposed electrode B can be successfully used to determine silver content in real samples.
Comparison of silver content in waste fixing solution and electroplating wastewater determined by the proposed Ag+-ISE and AAS
PPT Slide
Lager Image
aAverage of four replicate measurements±standard deviation (SD).
CONCLUSION
In this work, two amine phenol ligands, N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane (L1), N,N’-bis(2-hydroxybenzyl)-2,2-dimethyl-1,3-diaminopropane (L2), were used as the sensing carriers to construct ISEs, and a Ag + -ISE based on L1 was successfully prepared and characterized. The experimental observations indicated that the resulting electrode B incorporating L1 exhibited preferential potential recognition process and near-Nernst potential analytical performances toward Ag + in a wide linear range with a low detection limit. The proposed electrode could be used for silver assay in presence of Cu 2+ ion at 10-fold concentration at least. The excellent potential response characteristics such as near-Nernst sensitivity and high selectivity to Ag + over other tested cations could lead to the satisfactory measurement capabilities of silver concentration analysis in medicinal samples, environmental samples and etc.. Thus, the proposed Ag + -ISE based on L1 would be a promising and an alternative substitute for the silver assay in actual samples.
Acknowledgements
The authors would like to thank the PhD Foundation Project of Southwest University (No. SWU109026) and the National Nature Science Foundation of China (No. 20675064) to support this work.
References
Buhlmann P. , Pretsch E. , Bakker E. 1998 Chem. Rev. 98 1593 -    DOI : 10.1021/cr970113+
Oliveira I. A. M. , Risco D. , Vocanson F. , Crespo E. , Teixidor F. , Zine N. , Bausells J. , Samitier J. , Errachid A. 2008 Sens. Actu. B 130 295 -
Kim S. , Kim H. , Noh K. H. , Lee S. H. , Kim S. K. , Kim J. S. 2003 Talanta 61 709 -    DOI : 10.1016/S0039-9140(03)00329-1
Hassan A. Z. , Azizollah N. , Malyheh S. 2008 Anal. Lett. 41 2727 -    DOI : 10.1080/00032710802363560
Gupta V. K. , Singh A. K. , Gupta B. 2007 Anal. Chim. Acta 583 340 -    DOI : 10.1016/j.aca.2006.10.039
Mohammad R. A. , Hassan A. Z. 2008 Anal. Lett. 41 2251 -    DOI : 10.1080/00032710802238077
Mahajan R. K. , Kaur R. , Bhalla V. , Kumar M. , Hattori T. , Miyano S. 2008 Sens. Actu. B 130 290 -    DOI : 10.1016/j.snb.2007.08.006
Susan S. , Moslem J. 2009 Anal. Lett. 42 2026 -    DOI : 10.1080/00032710903083877
Ma Y. H. , Yuan R. , Chai Y. Q. , Wu X. , Zhou W. , Liu X. L. , Deng F. 2009 Anal. Lett. 42 2411 -    DOI : 10.1080/00032710903061147
Ion I. , Culetu A. , Costa J. , Luca C. , Ion A. C. 2010 Desalination 259 38 -    DOI : 10.1016/j.desal.2010.04.038
Siswanta D. , Nagatsuka K. , Yamada H. , Kumakura K. , Hisamoto H. , Shichi Y. , Toshima K. , Suzuki K. 1996 Anal. Chem. 68 4166 -    DOI : 10.1021/ac960396q
Shamsipur M. , Javanbakht M. , Lippolis V. , Garau A. , Flippo G. D. , Ganjali M. R. , Mousavi M. F. , Yari A. 2002 Anal. Chim. Acta 462 225 -    DOI : 10.1016/S0003-2670(02)00374-4
Gupta V. K. , Singh A. K. , Pal M. K. 2009 Anal. Chim. Acta 631 161 -    DOI : 10.1016/j.aca.2008.10.033
O’Connor K. M. , Henderson W. , O’Neill E. , Arrigan D. W. M. , Harris S. J. , Mckervey M. A. , Svehla G. 1997 Electroanalysis 9 311 -    DOI : 10.1002/elan.1140090410
Chen L. X. , Zeng X. S. , Ju H. F. , He X. W. , Zhang Z. Z. 2000 Microchem. J. 65 129 -    DOI : 10.1016/S0026-265X(00)00038-2
Chen L. X. , Ju H. F. , Zeng X. S. , He X. W. , Zhang Z. Z. 2001 Anal. Chim. Acta 437 191 -    DOI : 10.1016/S0003-2670(01)00994-1
Mahajan R. K. , Kaur I. , Kumar M. 2003 Sens. Actu. B 91 26 -    DOI : 10.1016/S0925-4005(03)00062-5
Singh A. K. , Saxena P. 2006 Anal. Bioanal. Chem. 385 90 -    DOI : 10.1007/s00216-006-0385-6
Mittal S. K. , Kumar A. S. K. , Kaur S. , Kumar S. 2007 Sens. Actu. B 121 86 -    DOI : 10.1016/j.snb.2006.04.001
Yan Z. N. , Lu Y. Q. , Li X. 2007 Sens. Actu. B 122 174 -    DOI : 10.1016/j.snb.2006.05.020
Mashhadizadeh M. H. , Taheri E. P. , Sheikhshoaie I. 2007 Talanta 72 1088 -    DOI : 10.1016/j.talanta.2007.01.021
Jeong T. , Lee H. K. , Jeong D. , Jeon S. 2005 Talanta 65 543 -    DOI : 10.1016/j.talanta.2004.07.016
Zamani H. A. , Ganjali M. R. , Salavati-Niasari M. 2008 Tran. Metal Chem. 33 995 -    DOI : 10.1007/s11243-008-9142-3
Singh A. K. , Gupta V. K. , Gupta B. 2007 Anal. Chim. Acta 585 171 -    DOI : 10.1016/j.aca.2006.11.074
Gupta V. K. , Khayat M. A. , Singh A. K. , Pal M. K. 2009 Anal. Chim. Acta 634 36 -    DOI : 10.1016/j.aca.2008.11.065
Pillai M. R. A. , Lo J. J. M. , Schlemper E. O. , Troutner D. E. 1990 Inorg. Chem. 29 1850 -    DOI : 10.1021/ic00335a018
Horning E. C. 1955 Organic Synthesis Colloids John Wiley & Sons New York, U.S.A. Vol. 3 140 - 141
Moody G. J. , Oke R. B. , Thomas J. D. R. 1970 Analyst 95 910 -    DOI : 10.1039/an9709500910
Buck R. P. , Lindner E. 1994 Pure Appl. Chem. 66 2527 -    DOI : 10.1351/pac199466122527
Ganjali M. R. , Akbar V. , Norouzi P. , Moghimi A. , Sepehrifard A. 2005 Eelectroanalysis 17 895 -    DOI : 10.1002/elan.200303169
Pan Z. T. (Wuhan University) 2006 Analytical Chemistry 5th Edn Higher Education Press Beijing, China 271 -
Zeng Y. H. , Lin S. C. 2004 Analytical Chemistry (Instrument Analysis Section) 2nd Edn Higher Education Press Beijing, China 277 - 279