Advanced
Determination of Copper Ion with 2-Mercaptobenzimidazol Immobilized on Surfactant-Coated Alumina
Determination of Copper Ion with 2-Mercaptobenzimidazol Immobilized on Surfactant-Coated Alumina
Journal of the Korean Chemical Society. 2007. Apr, 51(2): 141-146
Copyright © 2007, The Korean Chemical Society
  • Received : September 29, 2006
  • Published : April 20, 2007
Download
PDF
e-PUB
PubReader
PPT
Export by style
Article
Author
Metrics
Cited by
TagCloud
About the Authors
G. Absalan
A. Aghaei Goudi

Abstract
The present paper describes a procedure for separation, preconcentration and determination of trace amount of copper in natural water samples by using 2-mercaptobenzimidazol as the complexing agent. The proposed method is based on the surfactant aggregation formed on γ-alumina by mixing sodium dodecyl sulfate and γ-alumina in water; 2-mercaptobenzimidazol was incorporated into inner hydrophobic part of produced ad-micelles in acidic media to produce an assemble suitable for preconcentration and determination of copper ion. Optimum experimental conditions for adsorption of μg/ml levels of copper ions from aqueous solution by the adsorbent have been reported. The copper ions were quantitively adsorbed by the sorbent over the pH range of 7.1-8.0 and were quantitatively desorbed afterward by using sulfosalycilic acid as the eluent. The determination of copper was not interfered in the presence of common metal ions. The procedure was applied for analysis of river water sample. Relative standard deviation was found to be 4.91%.
Keywords
INTRODUCTION
Determination of heavy metal ions in natural waters is a task for analytical chemists in order to evaluate and interpret aquatic systems. However, in many cases the available analytical procedures do not show enough sensitivity thus many enrichment procedures have been proposed. The process involving extraction on a solid sorbent has received more acceptances due to a number of possible advantages including the availability of the solid sorbent, obtainment of large preconcentration factors, and facility for enrichment using systems with continuous flow. Besides, they usually do not need the use of organic solvents, which may be toxic. 1
Monitoring trace or ultra trace copper in natural, environment, biological and other real samples is very important. It is an essential element not only for life in mammals but also for plants and lower forms of organisms. However, it is toxic above a certain concentration level. Preconcentration procedures for copper determination 2 - 13 have been developed that involve different analytical techniques. Among them are liquid-liquid extractions using dithiocarbamate 2 and trioctylmethylammonium chloride 3 as complexing agents, copreciptation with magnesium hydroxide as collector, 4 filtration using a membrane filter. 5 Several systems of adsorption that use solid sorbents such as polyurethane foam loaded with diethyldithiocarbamate, 6 activated carbon, 7 , 8 amberlit XAD resins, 9 , 10 and naphthalene 11 - 15 have also been reported. The possibility of chromosorb absorbents for the enrichment of trace copper ion from aqueous solution without using a chelating reagent has been also proposed. 16 Bati et al . used pb-piperazinedithiocarbamate complex as a new solid-sorbent for preconcentration and separation of copper from oil samples. 17 The potential and utility of alumina and admicellar sorbents in chemical speciation, by taking the case of copper (II) in water as example was investigated by Hiraide. 18 Converted copper into the water insoluble complex with the aid of ammonium pyrrolidinedithiocarbamate has allowed a rapid and reliable determination of traces of copper in river and seawater. 19
2-Mercaptobenzimidazol (MBID) is slightly soluble in water and soluble in alcohol. It reacts with a wide variety of soft metal ions to form insoluble complexes and has been used as an extraction photometric reagent for Bi, Hg and Se. 20 Bagheri et al . 21 reported determination of both Fe 2+ and Fe 3+ species in water samples by selective retention of Fe 2+ complex on a silica gel column loaded with MBID. To the best of our knowledge surfactantcoated alumina has not been used as an adsorbent for the separation and preconcentration of metal ions from aqueous solutions. Surfactant molecules such as sodium dodecyl sulfate (SDS) are able to form self-aggregate structure called “hemi-micelles” or “ad-micelles” on solid surfaces such as alumina. It gives them a unique ability to immobilize metal complexing agents. Their interior tends to incorporate sparingly soluble metal complexing agents. 22 In this research, MBID was incorporated into inner hydrophobic part of produced ad-micelles when mixture of MBID, SDS, and γ-alumina (Al 2 O 3 ) particles was acidified to produce a sorbent (abbreviated as MBID/SDS/Al 2 O 3 in this text) suitable for preconcentration and determination of copper ion.
EXPERIMENTAL
- Apparatus
A flame atomic absorption spectrophotometer, FAAS, (PHILIPS PYE UNICAM SP9) with a deuterium lamp background correction was used. A copper hollow-cathode-lamp (VARIAN TECHTRON) was used with an operating current of 5 mA. Wavelength and spectral bandwidth were 324.7 and 0.2 nm, respectively. pH measurements were made with a Metrohm pH meter (model 632) using a combined glass electrode. Vacuum pump was Edwards 2.
- Reagents
All reagents were of analytical grades and were used without further purification. Standard copper ion solutions were prepared daily from acidified 1000 μg/ml copper (II) nitrate solution (Fluka). Other reagents were SDS (Fluka), nitric acid (Merck), sulfosalycilic acid (BDH), 2-mercaptobenzimidazol (Merck) and γ-alumina (Merck). Triply distilled water was used throughout the work. A MBID/SDS solution was prepared by dissolving 1 g SDS and 150 mg of MBID in 40 ml of 0.3 M aqueous ammonia diluting to 100 ml with triply distilled water.
- Preparation of the sorbent
Ten milliliters of MBID/SDS solution was added into a beaker containing 1.5 g γ-alumina suspended in 50 ml water. The pH of the mixture was adjusted to a value of 2.0 by drop-wise addition of diluted HNO 3 solution. The mixed solution was shaken for duration of 15 minutes and the produced solid phase (MBID/SDS/AL 2 O 3 ) was filtered with a sintered-glass funnel using the vacuum pump. The product was dried in room temperature and was used as adsorbent for copper throughout the work.
- Recommended procedure
A glass column (10×100 mm) packed with 0.20 g MBID/SDS/Al 2 O 3 (with a height of 2 mm in the column) was used as the operational column. Aliquot of 5 ml copper ion solution (8 μg/ml) was passed through the column at a flow rate of 1.5 ml/min. The adsorbed copper ion was eluted with 5 ml of 0.15 M sulfosalycilic acid (adjusted at pH 1.5) at flow rate of 2 ml/min. Percent recovery of copper ion was measured by FAAS after analyzing the eluted copper.
RESULTS AND DISCUSSION
γ-Alumina has positively charged surfaces that can strongly adsorb a negatively charged surfactant such as SDS in highly acidic solutions. A concentration of 7×10 -3 M SDS, below its critical micellar concentration (8×10 -3 M), was used in order to produce admicelles that could entrap MBID molecules into its produced cores. Above its critical micellar concentration, the excess of SDS would form micelles in the aqueous solution, which were not adsorbed on alumina surfaces and consequently were not suitable for solid-phase extraction purposes.
- pH of sample solution
Looking at the chemical structure of MBID 20 shows that it could behave as an amphiprotic substance because it has two basic amines and one acidic –SH (with a pK a value of 9.97) sites thus its complex formation ability with metal ions is pH dependent. A series of 5 ml of 8 μg/ml copper ion solutions with different pH values were introduced into packed column. Concentrations of copper ions in the effluents were analyzed by FAAS. . 1 shows that copper ions are completely adsorbed in the pH range of 7.1-8.0. A pH of 7.5, in the middle of the optimum pH range, was selected for further studies to avoid any abrupt changes in adsorption that may occur due to any changes in pH of the sample. By the way, this could consequently affect the precision of the measurement.
PPT Slide
Lager Image
Effect of pH of the sample solution on percent adsorption of copper ion. Experimental conditions: source; 5 ml of 8 μg/ml copper ion solutions at different pH values, sorbent; 0.2 g MBID/SDS/Al2O3.
- Sample solution flow rate
Percent sorption of 40.0 μg copper ion solution (5 ml of 8 μg/ml) on the surface of 0.2 g of sorbent as a function of sample flow rate was studied. The results ( . 2 ) showed that at flow rates less than 1.5 ml/min, copper ions were quantitatively adsorbed as copper ions and active sites of the sorbent have enough time for complex formation. Generally, the flow rate of the sample solution is a measure of contact time between the sorbent and copper ion; the lower the flow rate the higher will be the contact time. A flow rate of 1.0 ml/min was chosen for further studies in order to achieve a good precision. Furthermore, the chosen sample flow rate still provided quantitative adsorption with the additional benefit of a decrease in preconcentration time.
PPT Slide
Lager Image
Effect of sample flow rate on percent adsorption of copper ion. Experimental conditions: source; 5 ml of 8 μg/ml copper ion at pH 7, sorbent; 0.2 g MBID/SDS/Al2O3, eluent; 5 ml of 0.15M soulfosalycilic acid solution at pH 1.5.
- Eluent concentration
For desorbing 40.0 μg copper ion, adsorbed on 0.2 g sorbent, different concentrations of the eluent (sulfosalycilic acid) have been used. Percent copper ion recovered is shown as a function of sulfosalycilic acid concentration in . 3 . As it is shown in this figure, at concentrations more than 0.1 M of sulfosalycilic acid, copper ion was quantitatively desorbed from the sorbent surface. A concentration of 0.15 M of sulfosalycilic acid was selected for further studies.
PPT Slide
Lager Image
Effect of eluent concentration on percent recovery of adsorbed copper ion. Experimental conditions: source; 5 ml of 8 μg/ml copper ion solution at pH 7.0, sorbent; 0.2 g MBID/SDS/Al2O3, eluent; 5 ml of soulfosalycilic acid at different concentration.
- pH of the eluent solution
To find the optimum pH of the eluent solution for desorbing 40.0 μg of copper ion from sorbent surfaces, a series of eluent solutions with different pH values were used. The results are shown in . 4 . As it is seen, the percent recoveries of copper ions at pH ≤ 2 were almost quantitative. A pH of 1.5 for sulfosalycilic acid solution was selected for further research. In another experiment nitric acid solution at pH 1.5 was used as eluent. It was revealed that about 50% of copper ions were recovered. This showed that desorption of copper ion with sulfosalycilic acid is due to complex formation between sulfosalycilic acid and copper ion.
PPT Slide
Lager Image
Effect of pH of the eluent solution on percent recovery of adsorbed copper ion. Experimental conditions: source; 5 ml of 8 μg/ml copper ion solution at pH 7, sorbent; 0.2 g of MBID/SDS/Al2O3, eluent; 5ml of 0.15M soulfosalycilic acid with a flow rate of 2 ml/min at different pH values.
- Eluent flow rate
A series of copper ion (40.0 μg) sorbed on sorbent surfaces were eluted at different flow rates of sulfosalycilic acid solution. The percent recovery of copper ion as a function of eluent flow rate is shown in . 5 . The results showed that copper ion was desorbed almost quantitatively when eluent flow rate was less than 2.4 ml/min. An eluent flow rate of 2 ml/min was considered as the optimum flow rate because it would provide a quantitative recovery with a decrease in analysis time.
PPT Slide
Lager Image
Effect of eluent flow rate on percent recovery of adsorbed copper ion. Experimental conditions: source; 5 ml of 8 μg/ml copper ion solution at pH 7, sorbent; 0.2 g of MBID/SDS/Al2O3, eluent; 5 ml of 0.15M soulfosalycilic acid solution at pH 1.5.
- Effect of foreign ions on separation of copper ions
The effect of foreign ions present in 10 ml of a binary solution containing 8 μg/ml copper ion with different concentrations of the selected foreign ion, on recovery of copper ion were investigated. Adsorbed copper ions were determined using FAAS after eluting by sulfosalycilic acid. The results shown in 1 revealed that there was no serious interference in recovery of copper ion from any of the selected binary solutions.
Percent recovery of 80 ng/ml copper ion by sulfosalycilic acid in the presence of foreign ions. Experimental conditions: source, 10ml of 8 μg/ml copper ion containing foreign ions passed with a flow rate of 1.5 ml/min at pH 7.0 through the column; sorbent; 0.6 g of MBID/SDS/Al2O3; eluent, 10 ml of 0.15 M sulfosalysilic acid with a flow rate of 2 ml/min at pH 1.5
PPT Slide
Lager Image
Percent recovery of 80 ng/ml copper ion by sulfosalycilic acid in the presence of foreign ions. Experimental conditions: source, 10ml of 8 μg/ml copper ion containing foreign ions passed with a flow rate of 1.5 ml/min at pH 7.0 through the column; sorbent; 0.6 g of MBID/SDS/Al2O3; eluent, 10 ml of 0.15 M sulfosalysilic acid with a flow rate of 2 ml/min at pH 1.5
- Selectivity of the method
The ability of the system for separation of foreign ions, under optimum experimental conditions obtained for copper ion is referred as separation selectivity. A multinary solution containing 200 μg/ml each of Ni 2+ , Fe 3+ , Co 2+ , Zn 2+ , Cd 2+ , Al 3+ , Cr 2+ , Mg 2+ , Pb 2+ , Ca 2+ , Cs + and 8 μg/ml of copper ion were prepared. Aliquots of 200 ml of the multinary solution (source solution) were applied into the packed column. After passing through the column, the source solution was analyzed for each cation. Sulfosalycilic acid solution was applied afterward for eluting the possible adsorbed cations. Results, shown in 2 , revealed that the percent eluted of foreign ions made no serious interference in recovery of copper ion at the optimum experimental conditions applied for sorption/desorption of copper ion.
Selectivity of the method (percent recovery of foreign ions). Experimental conditions: source, 200 ml of selected ion solution passed through the column with a flow rate of 1.5 ml/min at pH 7.0; sorbent, 0.6 g of MBID/SDS/Al2O3; eluent, 2 ml of 0.15 M sulfosalysilic acid with a flow rate of 2 ml/min at pH 1.5
PPT Slide
Lager Image
Selectivity of the method (percent recovery of foreign ions). Experimental conditions: source, 200 ml of selected ion solution passed through the column with a flow rate of 1.5 ml/min at pH 7.0; sorbent, 0.6 g of MBID/SDS/Al2O3; eluent, 2 ml of 0.15 M sulfosalysilic acid with a flow rate of 2 ml/min at pH 1.5
- Analysis of real samples
In order to assess the applicability of the method to the analysis of real samples, recovery of spiked copper ions from Coar River water (Fars province, Iran) was studied. The results along with the quality of Coar River water are shown in 3 .
Replicate analysis of copper ion spiked in Coar River water (with a pH of 7.5, total hardness of 470 mg/l and a conductivity of 24.4 mS). Experimental conditions: source; 200 ml river water, sorbent; 2.0 g of MBID /SDS/Al2O3, eluent; 2 ml of 0.15M sulfosalycilic acid with a flow rate of 2 ml/min at pH 1.5
PPT Slide
Lager Image
Replicate analysis of copper ion spiked in Coar River water (with a pH of 7.5, total hardness of 470 mg/l and a conductivity of 24.4 mS). Experimental conditions: source; 200 ml river water, sorbent; 2.0 g of MBID /SDS/Al2O3, eluent; 2 ml of 0.15M sulfosalycilic acid with a flow rate of 2 ml/min at pH 1.5
CONCLUSIONS
The proposed method has advantages of good selectivity and simplicity, as there is no need for an additional elution step. It can selectively separate copper ion from other metal ions of much higher concentrations. The method can be successfully applied to the separation and determination of copper in river water samples. The R.S.D. of method was found to be 4.91%.
Acknowledgements
The authors are grateful to the Shiraz University Research Council for financial support of the work.
References
Christian G. D. 1994 Analytical chemistry Jhon Wiley & Sons, INC New York
Sachsenberg S. , Klenke T. , Krumbein W. E. , Zeeck E. 1992 Fresenius J. Anal. Chem. 342 163 -    DOI : 10.1007/BF00321714
Zhang X. Y. , Keiichi S. , Satoh S. A. , Sawada K. , Suzuki T. 1997 Anal. Sci. 13 891 -    DOI : 10.2116/analsci.13.891
Wu J. , Boyle E. A. 1997 Anal. Chem. 69 2464 -    DOI : 10.1021/ac961204u
Itoh J. L. , Miyake T. , Komata M. 1996 Nipp on Kagaku Kaishi 7 645 -    DOI : 10.1246/nikkashi.1996.645
Atanasova D. , Stefanova V. , Russeva E. 1998 Talanta 45 857 -    DOI : 10.1016/S0039-9140(97)00175-6
Santelli R. E. , Gallego M. , Valcarcel M. 1994 Talanta 41 817 -    DOI : 10.1016/0039-9140(94)80055-3
Uzawa A. , Narukawa T. , Okutani T. 1998 Anal. Sci. 14 395 -    DOI : 10.2116/analsci.14.395
Elci L. , Soylak M. , Dogan M. 1992 Fresenius J. Anal. Chem. 342 175 -    DOI : 10.1007/BF00321717
Jain V. K. , Sait S. S. , Shrivastav P. , Agrawal Y. K. 1997 Talanta 45 397 -    DOI : 10.1016/S0039-9140(97)00141-0
Taher M. A. 2001 J. Anal. Chem. 56 149 -    DOI : 10.1023/A:1009498703317
Taher M. A. , Puri B. K. 2000 Indian J. Chem. Tech. 7 168 -
Taher M. A. 2003 Microchim. Acta 141 101 -    DOI : 10.1007/s00604-002-0932-6
Taher M. A. 2003 Asian J. Chem. 15 1266 -
Taher M. A. , Bansal H. , Puri B. K. 2001 Anal. di Chimica 91 649 -
Cui Y. , Jiang G. , Lin J. , He B. 2002 Anal. Sci. 18 705 -    DOI : 10.2116/analsci.18.705
Bati B. , Cesur H. 2002 Anal. Sci. 18 1273 -    DOI : 10.2116/analsci.18.1273
Hiraide M. , Ishilcawa A. 2002 Anal. Sci. 18 199 -    DOI : 10.2116/analsci.18.199
Hiraide M. , Itoh T. 2004 Anal. Sci. 20 231 -    DOI : 10.2116/analsci.20.231
Cheng K. L. , Ueno K. , Imamura T. 1982 Handbook of Organic Analytical Reagents, Vol. 365 CRC Press Inc. 484 -
Bagheri A. , Gholami A. , Najafi A. 2000 Anal. Chim. Acta 424 (2) 233 -    DOI : 10.1016/S0003-2670(00)01151-X
Absalan G. , Ayatollahi Mehrdjardi M. 2003 Separation and Purification Technology 33 95 -    DOI : 10.1016/S1383-5866(03)00003-0