Advanced
Performance Evaluation of Hazardous Substances using Measurement Vehicle of Field Mode through Emergency Response of Chemical Incidents
Performance Evaluation of Hazardous Substances using Measurement Vehicle of Field Mode through Emergency Response of Chemical Incidents
Korean Journal of Environmental Agriculture. 2015. Dec, 34(4): 294-302
Copyright © 2015, The Korean Society of Environmental Agriculture
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
  • Received : March 06, 2015
  • Accepted : September 30, 2015
  • Published : December 31, 2015
Download
PDF
e-PUB
PubReader
PPT
Export by style
Share
Article
Author
Metrics
Cited by
About the Authors
Yeon-Hee Lee
Department of Environmental Engineering, University of Seoul, Seoul 02504, Korea
Seung-Ryul Hwang
Division of Research Development and Education, National Institute of Chemical Safety, Ministry of Environment, Daejeon 34111, Korea
Jae-Young Kim
Department of Agricultural Chemistry, Chungbuk National University, Cheongju 28644, Korea
Kyun Kim
Division of Research Development and Education, National Institute of Chemical Safety, Ministry of Environment, Daejeon 34111, Korea
Ji Hyun Kwak
Division of Research Development and Education, National Institute of Chemical Safety, Ministry of Environment, Daejeon 34111, Korea
Min Sun Kim
Division of Research Development and Education, National Institute of Chemical Safety, Ministry of Environment, Daejeon 34111, Korea
Joong Don Park
Division of Research Development and Education, National Institute of Chemical Safety, Ministry of Environment, Daejeon 34111, Korea
Junho Jeon
Department of Environmental Engineering, College of Engineering, Changwon National University, Changwon 51140, Korea
Ki Joon Kim
Yeongsan River Basin Environmental Office, Ministry of Environment, Gwangju 61945, Korea
Jin Hwan Lee
Division of Research Development and Education, National Institute of Chemical Safety, Ministry of Environment, Daejeon 34111, Korea
schem72@korea.kr

Abstract
BACKGROUND:
Chemical accidents have increased owing to chemical usage, human error and technical failures during the last decades. Many countries have organized supervisory authorities in charge of enforcing related rules and regulations to prevent chemical accidents. A very important part in chemical accidents has been coping with comprehensive first aid tool. Therefore, the present research has provided information with the initial applications concern to the rapid analysis of hazardous material using instruments in vehicle of field mode after chemical accidents.
METHODS AND RESULTS:
Mobile measurement vehicle was manufactured to obtain information regarding field assessments of chemical accidents. This vehicle was equipped with four instruments including gas chromatography with mass spectrometry (GC/MS), Fourier Transform Infrared Spectroscopy (FT-IR), Ion Chromatography (IC), and UV/Vis spectrometer (UV) to analyses of accident preparedness substances, volatile compounds, and organic gases. Moreover, this work was the first examined the evaluation of applicability for analysis instruments using 20 chemicals in various accident preparedness substances (GC/MS; 6 chemicals, FT-IR; 2 chemicals, IC; 11 chemicals, and UV; 1 chemical) and their calibration curves were obtained with high linearity ( r 2 > 0.991). Our results were observed the advantage of the high chromatographic peak capacity, fast analysis, and good sensitivity as well as resolution.
CONCLUSION:
When chemical accidents are occurred, the posted measurement vehicle may be utilized as tool an effective for qualitative and quantitative information in the scene of an accident owing to the rapid analysis of hazardous material.
Keywords
Introduction
With the remarkable development of Korea’s economy, chemicals have essential materials for agriculture, industry, national productivity and people’s daily life (Chung and Kim, 2009) . Specifically, the chemical industry has become one of the important industries concern to the rapid growth of Korean economy life (Chung and Kim, 2009) . However, recent surveys have emphasized that chemical materials bring great risks to human health, environment, and safety through chemical accidents or incidents (Fanelli ., 1980 ; US-EPA, 1993; Bertazzi ., 1998 ; Broughton, 2005) . Especially, these substances may cause infection to various diseases, if people are exposed for a long time, even at very low concentrations (Gupta and Verma, 2002) . Several countries focused on the chemical safety to strengthen national capabilities through the process of storage, transportation, production, usage, and disposal due to potential hazards during many years (Regulation (EC), 2006; Shi ., 2014) . Moreover, they have established supervisory authorities in charge of enforcing related rules and regulations to prevent chemical accidents (Rosenthal ., 2006 ; Wood ., 2008) . Nevertheless, chemical accidents and incidents have increased owing to human error and technical failures during the last decades (Krausmann ., 2011 ; Krausmann and Cruz, 2013 ; Zio and Aven, 2013) . For instance, major accidents including Bhopal (India, 1984), Seveso (Italy, 1976), Schweizerhalle (Switzerland, 1986), Toulouse (France, 2001), and Guangxi (China, 2008) were well known due to serious human health and environmental impacts. The chemical accident of Korea (Sep. 27, 2012), spillage of highly toxic hydrofluoric acid was also widely known because of various problems concern to health of local citizens and contamination of crops as well as other plants (Lim and Lee, 2012 ; Na ., 2013 ; An ., 2015) . In other words, this accident was released about 8 tons for transfer from a transporting vehicle to a storage tank through worker’s carelessness (Na ., 2013 ; An ., 2015) . The gaseous hydrogen fluoride was proliferated into the atmosphere instantly after the accident and five workers died as well as many local citizens suffered from respiratory irritation problems (Park, 2013 ; An ., 2015) . Furthermore, Korean government declared the area ‘special disaster zone’ and discarded approximately 9,000 tons of crops and plants (Park, 2013 ; An ., 2015) . Up to data, under the increasing the use of chemicals, chemical accidents have increased in Korea (2009 : 16, 2010 : 15, 2011 : 12, 2012 : 9, 2013 : 87, and 2014 : 104 cases). On the basis of the above considerations, Korean government suggested safety measures in relation to chemical accident (Chung and Kim, 2009 ; Shin, 2013) . Among numerous measures, the National Institute of Chemical Safety was distributed to mobile measurement vehicle at the six Local Environment Offices with the supports of the Ministry of Environment to field assessments of chemical accidents. This vehicle was designed to obtain information about human health and environmental contamination as the determination of concentration levels of chemicals (focusing on accident preparedness substances having high probability of chemical accident) through on-site analysis.
In the current research, we were the first to organize mobile measurement vehicle together with technical facilities including analysis instrument and sampling as well as preprocessing equipment. Moreover, we evaluated the various accident preparedness substances using analysis instruments as gas chromatography with mass spectrometry (GC/MS), Fourier Transform Infrared Spectroscopy (FT-IR), Ion Chromatography (IC), and UV-Vis spectroscopy in this vehicle.
Materials and methods
- Materials and chemicals
Vehicle manufacture was performed from Shin Kwang Technology, Gyeonggi, Korea during 2014. Analytical grade benzene, toluene, ethylbenzene, tert -butyl methyl ether, and m,p,o -xylenes were purchased from RESTEK (Bellefonte, PA, USA). Moreover, analytical grade benzyl chloride, m -cresol, p -nitrotoluene, and phenol were obtained from AccuStandard (New Haven, CT, USA). Multi-component Anion Mix 4* and Cation Mix 2* ion chromatography standards were purchased from AccuStandard (New Haven, CT, USA). The NH 4 OH were obtained from Sigma Chemical Co. (St. Louis, MO, USA). Other chemicals and solvents used in the present research were of analytical grade.
- Instrument
Mobile measurement vehicle was equipped with four instruments including GC/MS, FT-IR, IC, and UV. The GC/MS analysis was measured by a Griffin 460 (Griffin Analytical Technologies, LLC, USA) with a cylindrical ion trap (CIT) source. Secondly, the gas-phase FT-IR instrument (MIDAC Corporation, Westfield, USA) was recorded in MIDAC spectrometer in the range of 4500-800 cm -1 . Ion chromatography was accomplished by Thermo Scientific Dionex ICS-2100 (Sunnyvale, CA, USA) configured with a conventional analytical IC system. UV-Vis absorption was measured on a PerkinElmer UV WinLab spectrophotometer (PerkinElmer, Lambda, Germany).
- Analysis conditions of four instruments
The six chemical materials were analyzed with a Griffin 460 GC/MS with coupled to a cylindrical ion trap, equipped with a 15 m × 0.18 mm × 0.18 μm DM-5MS capillary column. The flow rate of the helium (99.999%) carrier gas was 1.0 ml/min. The injector and detector temperature were 200℃. The column oven temperature was programmed at 40℃ for 3 min, and then increased to 60℃ at 10℃/min and then increased to 250℃ at 70℃/min and held for 1 min. The remaining GC and mass conditions are shown in Table 1 .
Conditions of GC/MS for six chemical materials.
PPT Slide
Lager Image
Conditions of GC/MS for six chemical materials.
The two chemical materials using gas-phase FT-IR was analyzed in the 4500-800 cm -1 range using a Hg source, Ge bolometer detector at 4.2 K and add windows. The gas atmosphere in the cell was nitrogen at a pressure of 15 psi at room temperature. The maximum temperature during the measurements was around 1200 K, reached at the end of the run. The spectrum was recorded at a resolution of 1 cm -1 using 120 scans and a 10 L/min gas sample was injected to analyze using diaphragm pump. The separation of IC instrument was achieved using AS18 (250 × 4 mm) and CS12 (250 × 4 mm) to anion and cation materials with 25 μl sample injection volume. The flow rate was 1 ml/min and the column temperature was maintained at 30℃. The mobile phase was composed of 30 mM KOH (anion) and 20 mM MSA (cation) in water. The remaining conditions were programmed as follows : Detector cell heater 35℃, suppressor type SRS, Current 75 mA (anion) 59 mA (cation), CR-TC On. The absorbance of UV-VIS spectroscopy in each cell was measured at 640 nm.
Results and discussion
- Vehicle manufacture
The mobile measurement vehicle for field assessments regarding chemical accidents was supported by the Ministry of Environment. This vehicle is designed to obtain information about environmental concentration and human health impact from the exposure of chemical substances in chemical accidents (Posthuma ., 2014) . Thus, the National Institute of Chemical Safety has organized mobile measurement vehicle by Shin Kwang Technology. As seen for Fig. 1 , mobile vehicle was composed of three rooms such as drive & control site, analysis site, and preprocessing site.
PPT Slide
Lager Image
(A) Appearance of the mobile measurement vehicle. (B) The internal structure of the mobile measurement vehicle.
Among them, analysis site comprised of four instruments including GC/MS, FT-IR, IC, and UV-Vis spectroscopy. In the present study, we evaluated 20 accident preparedness substances having high probability of chemical accidents using analysis equipment. Moreover, this work reported for the first time that development of mobile measurement vehicle regarding chemical accidents from the Ministry of Environment in Korea.
- Preparation of calibration curve
To analysis of six chemical materials using GC/MS, calibration curves were prepared by trap time (0.05, 0.2, and 0.5 min) control with three concentrations (17.5, 70, and 175 μg) (Yun and Minyan, 2014) . The gas-phase FT-IR was performed with qualitative analysis using two chemicals (Harrison and Clark, 1992) . For ion chromatography analysis, the standard stock solutions were prepared using Multi-component Anion Mix 4* and Cation Mix 2* Ion Chromatography standard solutions (each 100 ppm). Calibration curves of anion materials were made by diluting the stock solution with water to five concentrations (1, 5, 10, 15, and 20 μg/mL) (Siriraks and Stillian, 1993) . Cation materials were constructed using diluting the stock solution with water to four different concentrations including 1, 3, 5, and 7 μg/mL, respectively (Siriraks and Stillian, 1993) . Calibration curve using UV-VIS spectroscopy was made by four concentrations (0.2, 0.4, 0.8, and 1.6 μg/mL) (Mérian ., 2010) .
- Applicability of analysis instrument
The GC/MS analysis provided to be reliable for rapid separation and identification of volatile organic compounds one of the harmful chemicals (Dumont and Delmas, 2003 ; Kim ., 2013) . Thus, to investigate the applicability for chemical materials of GC/MS, we examined six different chemicals, including benzene, toluene, ethylbenzene, tert -butyl methyl ether, m,p -xylene and o -xylene. As presented in Fig. 2 A, six peaks were separated within 5 min by GC/MS analysis and their retention times were as follows : peak 1, tR = 1.0 min; peak 2, tR = 1.3 min; peak 3, tR = 2.1 min; peak 4, tR = 3.7 min; peak 5, tR = 3.9 min; peak 6, tR = 4.3 min.
PPT Slide
Lager Image
(A) Representative GC/MS chromatogram of six chemical materials. (B-G) Mass fragmentation patterns of six chemical materials, (B) tert-butyl methyl ether, (C) benzene, (D) toluene, (E) ethylbenzene, (F) m,p-xylene, (G) o-xylene.
Peak 1 was detected three ion peaks at m/z 55, m/z 57, and m/z 73. The major fragmentation showed at m/z 73 and the minor ions were observed at m/z 55 and m/z 57 ( Fig. 2 B). On the basis of the described fragmentation patterns and mass spectral library, peak 1 was identified as tert -butyl methyl ether. However, this peak was not detected exactly molecular weight [M] + in analysis values of the full mass scan chromatogram and chemical library values. This phenomenon suggests that the structural characteristic of tert -butyl methyl ether may be responsible for the ion fragmentation patterns of GC/MS analysis. The MS spectrum of peak 2 presented the fragment ions at m/z 78, m/z 51. As can be seen in Fig. 2 C, one major ion peak ( m/z 78) and one minor ion peak ( m/z 51) showed in the chromatogram. Based on the mass spectral chemical library, the structure of peak 2 was tentatively determined as benzene. The GC/MS profile of peak 3 exhibited an identical molecular ion at m/z 92 with three fragmentation ions at m/z 91, m/z 65, and m/z 63 ( Fig. 2 D). Peak 3 was confirmed be toluene by comparing the mass spectral library. Subsequently, the major fragmentation ions of full mass scan analysis of peaks 4-6 were detected at m/z 106 ( Fig. 2 E), m/z 106 ( Fig. 2 F), and m/z 106 ( Fig. 2 G), respectively. According to the various fragmentation ions and mass spectral library, peaks 4-6 were assigned as ethylbenzene, m,p -xylene and o -xylene. For calibration curves, the trap time was operated with 0.05, 0.2, and 0.5 min to obtain at three concentration sequences (17.5, 70, and 175 μg). Furthermore, the correlation coefficient ( r 2 ) of each curve was higher than 0.991. In other words, their values showed 0.998 ( tert -butyl methyl ether), 0.994 (benzene), 0.993 (toluene), 0.992 (ethylbenzene), 0.991 ( m,p -xylene), and 0.996 ( o -xylene), respectively.
In general, it is well established that gas-phase FT-IR technique has been unprecedented attention due to the analyses of hazardous air pollutants, greenhouse gases, and ozone-destroying chemicals (Sharpe ., 2004) . Therefore, for determining of applicability concern to the analysis of gas-phase FT-IR, we investigated peak ranges using the exactly concentration (500 ppm) of NH 3 and CO standards. The spectrum distribution of NH 3 gas is known to exhibit the major peaks and several minor peaks in the ranges of 1200-800 cm -1 through qualitative analysis using gas-phase FT-IR. In our results, NH 3 spectrum was observed with the ranges of 1200-800 cm -1 ( Fig. 3 A) and its pattern was similar to that of chemical library system ( Fig. 3 B). The gas-phase FT-IR spectrum of the CO is depicted in Fig. 3 C. This spectrum contains two weakly overlapping peaks (2220-2060 cm -1 ) that were resolved by deconvolution to determine the peak positions. In the magnified spectrum, many peaks appears at 2140-2060 and 2220-2145 cm -1 ( Fig. 3 D). According to the chemical library spectrometric investigation, the peaks of CO spectrum was similar to those of the analyzed spectrum. Briefly, the CO gas has been demonstrated in the range of 2220-2060 cm -1 by FT-IR spectroscopy.
PPT Slide
Lager Image
FT-IR spectrums concern to NH3 and CO chemical materials. (A) FT-IR spectrum of NH3, (B) The magnified spectrum of NH3 (1200-800 cm-1), (C) FT-IR spectrum of CO, (D) The magnified spectrum of CO (2220-2060 cm-1).
To gather information on applicability for mobile measurement vehicle, we evaluated anion and cation components by ion chromatography. This analytical method was usually carried out for the determination of ionic species with sensitivity, accuracy, and reproducibility at very low concentration (Morganti ., 2007) . In the current research, five anions (F - , Cl - , Br - , SO 4 2- , and NO 3 - ) and six cations (Li + , Na + , NH 4 + , K + , Mg 2+ , and Ca 2+ ) using standard solutions. A complete chromatographic separation of anion materials was reached in 10 min at AS18 (250 × 4 mm) column condition. The IC analysis exhibited rapid separation and good sensitivity. The distribution of the anion materials in multi-component anion Mix 4* Ion Chromatography standard was observed in the following order : F - > Cl - > Br - > SO 4 2- > NO 3 - . The peak retention times of five anion materials were follows : peak 1 (F - ), tR = 3.1 min; peak 2 (Cl - ), tR = 4.5 min; peak 3 (Br - ), tR = 7.1 min; peak 4 (SO 4 2- ), tR = 7.8 min; peak 5 (NO 3 - ), tR = 9.4 min ( Fig. 4 A). Moreover, mean areas ( n = 3) generated from the standard solution was plotted against concentration (1, 5, 10, 15, and 20 μg/mL) to establish the calibration equation. A high linearity ( r 2 > 0.994) was obtained for individual calibration curve. In other words, the coefficient of each anion material showed 0.994 (F - ), 0.999 (Cl - ), 0.999 (Br - ), 0.999 (SO 4 2- ), and 0.999 (NO 3 - ), respectively.
PPT Slide
Lager Image
IC chromatograms of anion and cation materials. (A) IC chromatogram of five anion materials, (B) IC chromatogram of six cation materials.
In order to measure cation materials, we investigated the separation and detection of multicomponent cation Mix 2* standard by Ion Chromatography analysis. As illustrated in Fig. 4 B, six peaks including major peak and minor peaks were detected within 13 min at the flow rate of 1 mL/min in CS12 (250 × 4 mm) column. The most major peak was identified as Li + and the minor peaks were characterized as Na + , NH 4 + , K + , Mg 2+ , and Ca 2+ in comparison with the retention times of cation material index. Briefly, the distribution of the cation materials was in the following order : Li + (peak 1, tR = 3.6 min) > Na + (peak 2, tR = 4.3 min) > NH 4 + (peak 3, tR = 4.9 min) > K + (peak 4, tR = 6.0 min) > Mg 2+ (peak 5, tR = 10.0 min) > Ca 2+ (peak 6, tR = 12.5 min) ( Fig. 4 B). Also, their calibration curves were made using four different concentrations such as 1, 3, 5, and 7 μg/mL. All calibration curves using mean areas ( n = 3) generated from the standard solution had coefficients of linear correlation r 2 > 0.995. Consequently, six cation materials were detected with 0.999 (Li + ), 0.999 (Na + ), 0.995 (NH 4 + ), 0.999 (K + ), 0.999 (Mg 2+ ), and 0.999 (Ca 2+ ), respectively.
Finally, the analysis of UV-Vis spectrometry could be conducted with accurate measurement of light absorption according to the molecular structure and concentration of chemical material at different wavelengths (Haiss ., 2007) . For the screening concern to UV-Vis spectrophotometer analysis, we examined the NH 4 absorption at 640 nm. The stock solutions of the standard were diluted with water to four different concentrations (0.2, 0.4, 0.8, and 1.6 μg/mL). Calibration curve was obtained with high linearity ( r 2 =1.000) by plotting the standard concentration as a function of the absorption value.
Based on the above considerations, the evaluation of applicability using analysis instrument in the mobile measurement vehicle may be used as an important source for on-site analysis from chemical accident.
Conclusion
Consequently, we introduced the mobile vehicle regarding field assessments of chemical accidents and chemical exposures. This vehicle was the first to design to obtain information about environmental and human health problems owing to exposure toxic chemical materials in the scene of the accident from the National Institute of Chemical Safety in the Ministry of Environment of Korea. Moreover, it has shown three main functions : one) sampling of chemical substances; two) preprocessing function for on-site analysis; three) scientifically interpreting the data through analysis instrument. We also evaluated the applicability of analysis instruments including GC/MS, FT-IR, IC, and UV-Vis spectroscopy from various chemical materials.
References
An J. , Lee H. A. , Lee J. , Yoon H. O. (2015) Fluorine distribution in soil in the vicinity of an accidental spillage of hydrofluoric acid in Korea. Chemosphere 119 577 - 582    DOI : 10.1016/j.chemosphere.2014.07.043
Bertazzi P. A. , Bernucci I. , Brambilla G. , Consonni D. , Pesatori A. C. (1998) The Seveso studies on early and long-term effects of dioxin exposure: a review. Environmental Health Perspectives 106 (Suppl 2) 625 -    DOI : 10.1289/ehp.98106625
Broughton E. (2005) The Bhopal disaster and its aftermath: a review Environmental Health    DOI : 10.1186/1476-069X-4-6
Chung S. T. , Kim K. I. (2009) Case studies of chemical incidents and emergency information service in Korea. Journal of Loss Prevention in the Process Industries 22 (4) 361 - 366    DOI : 10.1016/j.jlp.2009.01.006
Dumont E. , Delmas H. (2003) Mass transfer enhancement of gas absorption in oil-in-water systems: a review. Chemical Engineering and Processing: Process Intensification 42 (6) 419 - 438    DOI : 10.1016/S0255-2701(02)00067-3
Fanelli R. , Bertoni M. P. , Castelli M. G. , Chiabrando C. , Martelli G. P. , Noseda A. , Garattini S. , Binaghi C. , Marazza V. , Pezza F. (1980) 2, 3, 7, 8-Tetrachlorodibenzo-p-dioxin toxic effects and tissue levels in animals from the contaminated area of Seveso, Italy. Archives of Environmental Contamination and Toxicology 9 (5) 569 - 577    DOI : 10.1007/BF01056936
Haiss W. , Thanh N. T. , Aveyard J. , Fernig D. G. (2007) Determination of size and concentration of gold nanoparticles from UV-vis spectra. Analytical Chemistry 79 (11) 4215 - 4221    DOI : 10.1021/ac0702084
Harrison P. G. , Clark E. N. (1992) FT-IR study of the gas phase thermal decomposition of dimethyldivinyltin in the presence of oxygen. Journal of Organometallic Chemistry 437 (1) 145 - 156    DOI : 10.1016/0022-328X(92)83439-O
Gupta V. K. , Verma N. (2002) Removal of volatile organic compounds by cryogenic condensation followed by adsorption. Chemical Engineering Science 57 (14) 2679 - 2696    DOI : 10.1016/S0009-2509(02)00158-6
Kim K. , Lee J. , Lee S. , Hwang S. , Kim Y. , Seok G. (2013) Study on the analytical method using GC-MS for the accident preparedness substances. Analytical Science and Technology 26 (1) 80 - 85    DOI : 10.5806/AST.2013.26.1.080
Krausmann E. , Cozzani V. , Salzano E. , Renni E. (2011) Industrial accidents triggered by natural hazards: an emerging risk issue. Natural Hazards and Earth System Science 11 (3) 921 - 929    DOI : 10.5194/nhess-11-921-2011
Krausmann E. , Cruz A. M. (2013) Impact of the 11 March 2011, Great East Japan earthquake and tsunami on the chemical industry. Natural Hazards 67 (2) 811 - 828    DOI : 10.1007/s11069-013-0607-0
Lim H. S. , Lee K. (2012) Health care plan for hydrogen fluoride spill, Gumi, Korea. Journal of Korean medical science 27 (11) 1283 - 1284    DOI : 10.3346/jkms.2012.27.11.1283
Mérian T. , Debarnot D. , Rouessac V. , Poncin-Epaillard F. (2010) Ammonia absorption study of pulsed-plasma polyaniline by quartz crystal microgravimetry and UV/vis spectrometry. Talanta 81 (1) 602 - 608    DOI : 10.1016/j.talanta.2009.12.048
Morganti A. , Becagli S. , Castellano E. , Severi M. , Traversi R. , Udisti R. (2007) An improved flow analysis–ion chromatography method for determination of cationic and anionic species at trace levels in Antarctic ice cores. Analytica Chimica Acta 603 (2) 190 - 198    DOI : 10.1016/j.aca.2007.09.050
Na J. Y. , Woo K. H. , Yoon S. Y. , Cho S. Y. , Song I. U. , Kim J. A. , Kim J. S. (2013) Acute symptoms after community hydrogen fluoride spill. Annals of Occupational and Environmental Medicine 25 1 - 12    DOI : 10.1186/2052-4374-25-17
Park S. B. (2013) Alter over South Korea toxic leaks Nature 494 15 - 16    DOI : 10.1038/494015a
Posthuma L. , Wahlstrom E. , Nijenhuis R. , Dijkens C. , de Zwart D. , van de Meent D. , Hollander A. , Brand E. , den Hollander H. A. , van Middelaar J. , van Dijk S. , Hall E. F. , Hoffer S. (2014) The Flash environmental assessment tool: Worldwide first aid for chemical accidents response, pro action, prevention and preparedness. Environment International 72 140 - 156    DOI : 10.1016/j.envint.2014.04.018
Rosenthal I. , Kleindorfer P. R. , Elliott M. R. (2006) Predicting and confirming the effectiveness of systems for managing low‐probability chemical process risks. Process Safety Progress 25 (2) 135 - 155    DOI : 10.1002/prs.10124
Sharpe S. W. , Johnson T. J. , Sams R. L. , Chu P. M. , Rhoderick G. C. , Johnson P. A. (2004) Gas-phase databases for quantitative infrared spectroscopy. Applied Spectroscopy 58 (12) 1452 - 1461    DOI : 10.1366/0003702042641281
Shi S. , Cao J. , Feng L. , Liang W. , Zhang L. (2014) Construction of a technique plan repository and evaluation system based on AHP group decisionmaking for emergency treatment and disposal in chemical pollution accidents. Journal of Hazardous Materials 276 200 - 206    DOI : 10.1016/j.jhazmat.2014.05.034
Shin I. J. (2013) The effective control of major industrial accidents by the Major Industrial Accident Prevention Centers (MAPC) through the Process Safety Management (PSM) grading system in Korea. Journal of Loss Prevention in the Process Industries 26 (4) 803 - 814    DOI : 10.1016/j.jlp.2013.02.011
Siriraks A. , Stillian J. (1993) Determination of anions and cations in concentrated bases and acids by ion chromatography: Electrolytic sample pretreatment. Journal of Chromatography A 640 (1) 151 - 160    DOI : 10.1016/0021-9673(93)80177-A
Wood M. H. , Fabbri L. , Struckl M. (2008) Writing seveso II safety reports: New EU guidance reflecting 5 years’ experience with the directive. Journal of Hazardous Materials 157 (2) 230 - 236    DOI : 10.1016/j.jhazmat.2007.12.089
Yun Z. , Minyan L. (2014) The method study on emergency detection of aromatic compounds and chlorides based on portable GC-MS. Procedia Engineering 84 731 - 735    DOI : 10.1016/j.proeng.2014.10.489
Zio E. , Aven T. (2013) Industrial disasters: Extreme events, extremely rare. Some reflections on the treatment of uncertainties in the assessment of the associated risks. Process Safety and Environmental Protection 91 (1) 31 - 45    DOI : 10.1016/j.psep.2012.01.004