Optimization in Detecting Multiply-charged Protein Ions using MALDI TOF-MS
Optimization in Detecting Multiply-charged Protein Ions using MALDI TOF-MS
Mass Spectrometry Letters. 2013. Mar, 4(1): 21-23
Copyright ©2013, Korean Society Mass Spectrometry
  • Received : March 11, 2013
  • Accepted : March 20, 2013
  • Published : March 29, 2013
Download
PDF
e-PUB
PubReader
PPT
Export by style
Share
Article
Author
Metrics
Cited by
TagCloud
About the Authors
Jihyeon Lee
Jangmi Hong
Taehee Kim
Jeongkwon Kim
jkkim48105@cnu.ac.kr

Abstract
The effects of trifluoroacetic acid (TFA) were evaluated on the generation of multiply charged ions of cytochrome c in a 2-nitrophloroglucinol (2-NPG) matrix in high-vacuum, matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). The presence of 1% TFA in the 2-NPG matrix solution was more effective in generating multiply charged protein ions than matrix solutions containing 0.1% or 0% TFA. Regarding the matrix itself, with 1% TFA, 2-NPG was significantly more effective in generating multiply charged ions than 2,5-dihydroxybenzoic acid (2,5-DHB). The maximum charge state of cytochrome c was +8 when using a 2-NPG matrix containing 1% TFA.
Keywords
Introduction
Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) is a soft ionization technique frequently used for the analysis of proteins. MALDI matrices are effective energy mediators that facilitate analyte ionization. In sample preparations for MALDI-TOF MS, the analyte is dispersed in the matrix solution and deposited as a single sample spot. A pulsed laser beam then strikes the sample spot, causing desorption of both the matrix and analyte. MALDI-TOF MS typically provides a mass spectrum of mostly singly charged proteins in a vacuum system, while electrospray ionization MS typically yields multiply charged protein ion peaks. 1 , 2
Generally, conventional matrices such as 2,5-dihydro- xybenzoic acid (2,5-DHB), sinapic acid, and α-cyano-4-hydroxycinnamic acid generate mostly singly charged protein ions in high-vacuum MALDI systems. Recently, however, atmospheric pressure MALDI reportedly can produce multiply charged protein ions under certain conditions. 3 This technique is referred to as laser spray ionization (LSI). More recently, the employment of a new matrix, 2-nitrophloroglucinol (2-NPG), was shown to yield multiply charged protein ions even in high vacuum. 4 Since most commercial MALDI-TOF MS instruments operate in high vacuum, the production of multiply charged protein ions, facilitated by the 2-NPG matrix, would significantly reduce the m/z scan range and allow high mass accuracy analyses with enhanced sensitivity.
The current study investigated the efficiency of generating multiply charged protein ions under high-vacuum MALDI-TOF MS conditions using 2-NPG as a sample matrix, where various levels of trifluoroacetic acid (TFA) were added to the sample matrix to enhance and optimize the generation of multiply charged protein ions. This is the first systematic investigation on the effects of acid additives in a MALDI sample matrix on the generation of multiply charged protein ions.
Experimental
2,5-DHB, 2-NPG, TFA, acetonitrile (ACN), and horse heart cytochrome c were purchased from Sigma-Aldrich (St. Louis, MO, USA).
All mass spectra were obtained on a Voyager MALDI-TOF mass spectrometer (Applied Biosystems, Foster City, CA, USA) equipped with a nitrogen laser (337 nm, 3 ns pulse, 20 Hz repetition rate) in linear ion mode for a mass range of 500~15,000 Da.
To prepare the 2,5-DHB matrix solutions, 10 mg of 2,5-DHB was dissolved in 1 mL of aqueous 50% ACN/1% TFA. The 2-NPG matrix solution was prepared by dissolving10 mg of NPG into 1 mL of aqueous 50% ACN, 50% ACN/0.1% TFA, or 50% ACN/1% TFA solutions. To make a protein stock solution, 10 mg of cytochrome c was dissolvedin 1 mL of distilled water. The stock solution was then diluted with distilled water to make a 10 pmol/μL cytochrome c sample solution.
A dried-droplet sample preparation method was adopted for MALDI-TOF MS analyses. A 1-μL droplet of the mixturesolution (matrix:sample = 1:1 by volume) was placed on a MALDI plate and allowed to dry, forming a microcrystal layer.
Results and Discussion
Molecular structures of 2-NPG and 2,5-DHB are shown in Figure 1 . Note that 2-NPG is a novel matrix used to enhance the generation of multiply charged protein ions in high-vacuum MALDI-MS analysis. 2,5-DHB is one of the most commonly used matrix materials in MALDI-MS analyses. 5
Figure 2 shows MALDI mass spectra of cytochrome c using the 2-NPG and 2,5-DHB matrices. The theoretical average m/z values for the individual charge states are given
PPT Slide
Lager Image
The molecular structures of (a) 2-NPG and (b) 2,5-DHB.
PPT Slide
Lager Image
MALDI mass spectra of cytochrome c using (a) 2-NPG in 50% ACN, (b) 2-NPG in 50%ACN/0.1% TFA, (c) 2-NPG in 50% ACN/1% TFA, and (d) 2,5-DHB matrix in 50% ACN/1% TFA. Charge state of each protein peak is labeled.
The theoretical average m/z value for each charge state of cytochrome ca)a)The molar mass of cytochromecwas calculated by adding the mass of amino acid sequences from 2-105 of horse heart cytochromec(Swiss-Prot accession number/entry name = P00004/ CYC_HORSE) (C524H841N144O151S4), the mass of a heme group (C34H32O4N4Fe1), and the mass of N-terminal acetylation (C2H2O1).
PPT Slide
Lager Image
The theoretical average m/z value for each charge state of cytochrome ca) a)The molar mass of cytochrome c was calculated by adding the mass of amino acid sequences from 2-105 of horse heart cytochrome c (Swiss-Prot accession number/entry name = P00004/ CYC_HORSE) (C524H841N144O151S4), the mass of a heme group (C34H32O4N4Fe1), and the mass of N-terminal acetylation (C2H2O1).
in Table 1 . The MALDI mass spectrum of cytochrome c using the 2-NPG matrix without any TFA contains peaks representing three charge states: +1, +2, and +3 ( Figure 2 (a)). The presence of 0.1% TFA provided additional higher charge states up to +5 ( Figure 2 (b)). The presence of 1% TFAgenerated even higher charge states up to +8 ( Figure 2 (c)). For comparison, the MALDI mass spectrum of cytochrome c in the 2,5-DHB matrix with 1% TFA is shown in Figure 2 (d), where only a dominant singly charged ion peak and a minor doubly charged ion peak were observed.
Higher concentrations of TFA (> 1%) were also evaluated, but no improvements over the performance of matrices containing 1% TFA were observed. The 2-NPG matrix containing 1% TFA provided the highest abundance of multiply charged cytochrome c ions, yielding both the highest charge states and the highest signal-to-noise ratios.
Trimpin and coworkers 4 were the first to report the use of a 2-NPG matrix to enhance the generation of multiply charged protein ions in high-vacuum MALDI analyses. However, no mention was made as to the amount of acid, if any, added to the matrix solution. A 2-NPG matrix in 50:50 ACN:H 2 O was also used without acid in a recent high-vacuum application of LSI. 6 The current analysis demonstrates that the addition of a small amount of TFA (1.0%) to the 2-NPG matrix solution can significantly improve the detection of multiply charged protein ions.
Conclusion
The generation of multiply charged protein ions using a 2-NPG matrix containing TFA was evaluated and optimizedin a high-vacuum MALDI system. A 2-NPG matrix containing 1% TFA was optimal in terms of signal-to-noise and peak abundance for creating multiply charged protein ions. These results are a significant improvement over analogous analyses using a 2,5-DHB matrix. The generationof multiply charged protein ions shifts larger molecules into a lower m/z range and results in higher-quality mass spectra.
Acknowledgements
The authors acknowledge support for this work by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education, Science and Technology (2010-0010776).
References
Lomeli S. H. , Yin S. , Ogorzalek Loo R. R. , Loo J. A. 2009 J. Am. Soc. Mass Spectrom. 20 593 -    DOI : 10.1016/j.jasms.2008.11.013
Sterling H. J. , Daly M. P. , Feld G. K. , Thoren K. L. , Kintzer A. F. , Krantz B. A. , Williams E. R. 2010 J. Am. Soc. Mass Spectrom. 21 1762 -    DOI : 10.1016/j.jasms.2010.06.012
Trimpin S. , Inutan E. D. , Herath T. N. , McEwen C. N. 2010 Anal. Chem. 82 11 -    DOI : 10.1021/ac902066s
Trimpin S. , Ren Y. , Wang B. , Lietz C. B. , Richards A. L. , Marshall D. D. , Inutan E. D. 2011 Anal. Chem. 83 5469 -    DOI : 10.1021/ac2007976
Seo M. , Kim J. , Park S. , Lee J.-H. , Kim T. , Lee J. 2012 Mass Spectrom. Lett. 3 47 -    DOI : 10.5478/MSL.2012.3.2.47
Nyadong L. , Inutan E. D. , Wang X. , Hendrickson C. L. , Trimpin S. , Marshall A. G. 2013 J. Am. Soc. Mass Spectrom. 24 320 -    DOI : 10.1007/s13361-012-0545-1