Determination of Mercury in Fly Ash by Using Flow Injection Cold Vapor Isotope Dilution Inductively Coupled Plasma Mass Spectrometry
Determination of Mercury in Fly Ash by Using Flow Injection Cold Vapor Isotope Dilution Inductively Coupled Plasma Mass Spectrometry
Mass Spectrometry Letters. 2012. Jun, 3(2): 58-61
Copyright ©2012, Korean Society Mass Spectrometry
  • Received : February 17, 2012
  • Accepted : June 18, 2012
  • Published : June 28, 2012
Download
PDF
e-PUB
PubReader
PPT
Export by style
Share
Article
Author
Metrics
Cited by
TagCloud
About the Authors
Jung Ki Suh
Division of Metrology for Quality of Life, Analytical Chemistry Laboratory (KRISS), P.O. Box 102, Yusung, Daejon 305-600, Korea
jksuh@kriss.re.kr
Hyung Sik Min
Division of Metrology for Quality of Life, Analytical Chemistry Laboratory (KRISS), P.O. Box 102, Yusung, Daejon 305-600, Korea
Mohammad Kamruzzaman
Department of Chemistry, Kyungpook National University, Daegu 702-701, Korea
Sang Hak Lee
Department of Chemistry, Kyungpook National University, Daegu 702-701, Korea

Abstract
A method based on flow injection-isotope dilution-cold vapor-inductively coupled plasma mass spectrometry (FI-ID-CV-ICP/MS) has been applied to determine trace level of mercury in fly ash. 200 Hg isotopic spike was added to 0.25 g of BCR176R fly ash and then decomposed by microwave digestion procedure with acid mixture A (8 mL HNO 3 + 2 mL HCl + 2 mL HF) and acid mixture B (8 mL HNO 3 + 2 mL HClO 4 + 2 mL HF) for applying IDMS. Mercury cold vapor was generated by using reductant solution of 0.2% (w/w) NaBH 4 and 0.05% (w/w) NaOH. The measurements of n( 200 Hg)/n( 202 Hg) isotope ratio was made using a quadrupole ICP/MS system. The accuracy in this method was verified by the analysis of certified reference material (CRM) of fly ash (BCR 176R). The indicative value of Hg in BCR 176R fly ash was 1.60 ± 0.23 mg/kg (k = 2). The determined values of Hg in BCR 176R fly ash by the method of FI-CV-ID-ICP/MS described in this paper were 1.60 ± 0.24 mg/kg (k = 3.18) and the analysis results were in well agreement with the indicative value within the range of uncertainty.
Keywords
Introduction
Mercury (Hg) is a major concern among hazardous heavy metals because of its high volatility and high toxicity. In environmental matrices, Hg usually occurs in very low contents, so that highly sensitive analytical methods are needed for the determination of the element. 1 Therefore, with the increasing importance of the effect of mercury in environment on health, various analytical methods to detect a trace amount of mercury in various samples have been continuously developed. In atomic spectroscopy, analytical methods of mercury such as atomic absorption spectrometry (AAS), 2 atomic fluorescence spectroscopy (AFS), 3 and inductively coupled plasma mass spectrometry (ICP/MS) 4 have been reported. Especially, isotope dilution-inductively coupled plasma mass spectrometry (ID-ICP/MS) had been developed and used as a primary method in inorganic atomic analytical method because ICP/MS is applicable to analyze isotope analysis. 5 9 A method of isotope dilution-cold vapor-inductively coupled plasma mass spectrometry (ID-CV-ICP/MS) was developed in National Institute of Standards and Technology (NIST) at 2001 and applied to analyze small amount of mercury in various NIST Standard Reference Material (SRM) samples. 10 National Research Council of Canada (NRCC) which is the national standard research institute in Canada adopted the ID-CV-ICP/MS method to determine mercury in river standard samples. 11
In this study, an ID-CV-ICP/MS method coupled with flow injection (FI) has been applied to determine the content of Hg in BCR 176 fly ash CRM. Isotopic ratios were measured at two different signal processing mode in FI, that is, average mode and maximum mode, and the reproducibility of the obtained isotopic ratios were compared.
For mercury cold vapor generation, the solutions of 0.05% (w/w) NaOH and 0.2% (w/w) NaBH 4 with3 % (w/w) HCl solution were used as reductant. For isotope dilution method, the 200 Hg isotope was used as an enriched spike and 202 Hg was used as a reference isotope.
Two different acid mixtures (acid mixture-A: 8mL HNO 3 + 2mL HCl + 2mL HF, acid mixture-B: 8mL HNO 3 + 2mL HClO 4 + 2mL HF) were compared for fly acid sample digestion. The mineral acid mixture-A (HNO 3 + HCl + HF) was used according to EPA method 3052. The mineral acid mixture- B (HNO 3 + HClO 4 + HF) was used according to the method that has been used for decomposition of NIST 1633a coal fly ash. 12 To investigate the validity of the method described in this paper, Mercury in BCR CRM 176R fly CRM was analyzed and compared with the indicative value.
Experimental
- Instruments and reagents
Dynamic Reaction Cell-ICP-MS (Perkin-Elmer SCIEX Elan 6100) was used in this study and FIAS 400 (Perkin-Elmer) device was used for continuous flow injection. A 1000mg/kg (in 10% HNO 3 ) Hg standard solution was prepared and certified at the Korea Research Institute of Standards and Science (KRISS). BCR-176R fly ash was provided by the Institute for Reference Materials and Measurements (IRMM). Concentrated isotope of 200 Hg (abundance 84.45%) was purchased from TRACE SCIENCES (International Corp., Ontario, Canada). NaBH 4 (Sigma-Aldrich, granules, 99.99% Cat#480886), NaOH (Sigma-Aldrich, pellets, 99.99%, semiconductor grade, Catalog No. 30657-6), HCl (Dong Woo Fine Chem, Korea, Iksan, electronic grade) were used to produce mercury cold vapor. Milestone Ethos Plus/HPR 1000/100 Rotor (Italy) was used for microwave digestion.
- Flow injection-cold vapor-ICP/MS system
A flow injection device to produce hydride is shown in Figure 1 . HCl solution (3%, w/w) was used as carrier with 9-12mL/min flow rate and 0.2% NaBH 4 /0.05% NaOH as reductant with 5-7mL/min flow rate was used in this study. Containing volume of sample loop was 500 μ;L. Gaseous mercury from gas/liquid separator was directly injected into ICP/MS through PTFE tubing (1/8 inch × 1m). In this step, flow rate of argon gas that was injected from argon nebulizer gas line was 1.0mL/min.
- Sample decompositions
0.25 g BCR 176R fly ash was measured and taken into a Teflon reaction vessel of microwave digestion. Then some amount of concentrated isotope 200 Hg was added into the measured sample. And the sample was decomposed after adding acid mixed solution A (8 mL HNO 3 + 2 mL HCl + 2 mL HF) and acid mixed solution B (8 mL HNO 3 +2mL HClO 4 + 2 mL HF). In the microwave heating, the sample was decomposed at 120 ℃ for 5 min, 80 ℃ for 3 min, 180 ℃ for 10 min, 220 ℃ for 35 min, step by step.
Results
- Signal profiling data
Typical signal and data acquisition conditions using the FI-CV-ICP/MS system is shown in Figure 2 . This figure shows the three replicate signals for 1.0 ng/g of mercury solution. The introduction of mercury vapor into ICP-MS resulted in a single isotope sensitivity of 4.5 × 10 4 counts·s −1 /ng·g −1 for 202 Hg. The RSD of three replicates was within 0.5%. The reproducibility means significantly reduced memory compared with a traditional solution nebulization ICPMS. 13
- Repeatability for measurement of isotope ratio
In this system, isotope ratio n( 200 Hg/ 202 Hg) and intensity with eight times analysis for 102 min to investigate reproducibility of analysis of isotope ratio are shown in Table 1 . The maximum and average mode of peak processing mode were used to investigate reproducibility using approximately 1.0 ng/g concentrated mercury sample. The relative standard deviation (RSD) of the intensity of the isotope 200 Hg and 202 Hg was high about 9-10%, but the RSD of the isotope ration( 200 Hg / 202 Hg) was within the range of 0.65% at maximum mode and 0.85% at average mode
PPT Slide
Lager Image
Diagram of flow injection cold vapor generation system used in this experiment.
PPT Slide
Lager Image
Typical signal profile and data acquisition conditions for mercury isotopes recorded in FI-CV-ICP/MS system.
Repeatability of intensities and isotopic ratios of 200Hg and 202Hg in maximum peak processing mode of FI-CV-ICP/MS system (Concentration of Hg: 1.0 μg/kg).
PPT Slide
Lager Image
Repeatability of intensities and isotopic ratios of 200Hg and 202Hg in maximum peak processing mode of FI-CV-ICP/MS system (Concentration of Hg: 1.0 μg/kg).
based on the results.
- Determination of Hg in fly ash by FI-CV-ID-ICP/MS method
Each parameter of IDMS obtained from the following formulae is shown in Table 2 . 14
PPT Slide
Lager Image
Hg content in standard samples, BCR176R (Fly ash from a city waste incineration plant) was determined by the isotope dilution method using the FI-CV-ICP/MS device. To use an isotope dilution method, the blend solution (blend b) was firstly prepared by adding some amount of enriched spike isotope ( my ) into some amount of sample ( mx ) and then the isotope ratio ( Rb ) was determined. At the same time, the blend solution (blend b') was prepared by adding some amount of enriched spike isotope ( my' ) into primary assay standard ( mx ) and then the isotope ratio ( Rb' ) was determined.
Typical value of each parameters of IDMS equation and the analytical results of Hg in BCR-176R fly ash.
PPT Slide
Lager Image
Typical value of each parameters of IDMS equation and the analytical results of Hg in BCR-176R fly ash.
Analytical results of mercury in BCR-176R fly ash by FI-CV-ID-ICP/MS.
PPT Slide
Lager Image
Analytical results of mercury in BCR-176R fly ash by FI-CV-ID-ICP/MS.
Representative values of each parameter and calculated concentration values by equation ( Eq . 1) with isotope dilution method are shown in Table 3 . Here, the dilution factor (D) in BCR 176 fly ash was 0.0000593 because primary assay standard (KRISS) solution concentrated at 998.968 mg/kg was diluted to 0.059 mg/kg.
Dry mass determination of BCR176R fly ash sample was carried out by drying in ventilated oven at 105 ℃ for 3 hours, until constant weight is reached. Measured moisture content in
Summarize analytical results and calculation of combined and expanded uncertainty for Hg determination in BCR-176R by FI-CV-ID-ICP/MS.
PPT Slide
Lager Image
Summarize analytical results and calculation of combined and expanded uncertainty for Hg determination in BCR-176R by FI-CV-ID-ICP/MS.
0.25 g portions of BCR176R was 0.36% ± 0.02%, and the value of dry mass correction factor ( W ) was 0.9963. Mercury concentrations calculated by each representative values are 1.596 mg/kg at BCR-176R.
The analytical results of mercury in a BCR-176R fly ash by FI-CV-ID-ICP/MS are summarized in Table 3 . The indicative value of Hg in BCR 176R fly ash was 1.60 ± 0.23 mg/kg (k = 2). The determined values of Hg in BCR 176R fly ash by the method described in this paper were 1.60 ± 0.24 mg/kg (k = 3.18) and the analysis results were in well agreement with the indicative value within the range of uncertainty.
- Evaluation of measurement uncertainty
The typical example of the combined and expanded uncertainty calculation for Hg determination in BCR-176R by FI-CV-ID-ICP/MS are summarized in Table 4 . In Sub-sample-1, concentration of mercury is (1.591 ± 0.074) mg/kg. Here, the uncertainty value, ± 0.074, of measurement is systematic uncertainty. Four sub-samples produced similar systematic uncertainty values. Table 4 shows average value of four subsamples, valid degree of freedom ( ν ), pooled standard deviation, combind uncertainty (u c ), coverage factor (k), expanded uncertainty (U) of each samples by ISO guide. 15 , 16
- Linearity and limit of detection
Validation processes for the proposed method were conducted as recommended by an international guide. 17 In order to investigate the linearity and limit of detection for the proposed method, FI-ICP-MS signal intensities were examined as a function of mercury concentrations ranging from a 0.3 μ;g/kg to 1.11 μ;g/kg of 200 Hg and 202 Hg isotope. As shown in Figure 3 , the values of R 2 of 0.9997 and 0.9996 under the two different isotopes suggest very good linear relationships occurred between the ICP-MS signal intensities and the examined mercury concentration within the ranges.
The limit of detection (LOD) for Hg concentrations using the proposed analytical system was determined based on the following criterion. 18
PPT Slide
Lager Image
Where CDL = LOD expressed in concentration (μ;g/kg)
t (1-a, n-1) = Student’s t statistic at a specified confidence level and degree of freedom (dimensionless)
Sb = the standard deviation of the background computed from at least 10 replicate measurements (CPS; counts per second)
m = the slope of the analytical calibration curve (CPS per μ;g/kg).
The assigned value of the t statistics, t (1-a, n-1) , for 10 replicates ( n = 9) tested in this study is 2.26 at 95% confidence level. Based on 10 independent measurements for the background ion current, the obtained value for the standard deviation of background, Sb , was 70 cps. For the 200 Hg isotope, the slope of the calibration curves was 20047 cps per μ;g/kg. With these values, the minimum detectable concentration, CD , of an
PPT Slide
Lager Image
Calibration curve in FI-CV-ICP/MS system for Hg isotopes.
approximately 0.008 μ;g/kg was obtained. And, for the 202 Hg isotope, the slope of the calibration curves was 25476 cps per μ;g/ kg. With these values, the minimum detectable concentration, CD , of an approximately 0.006 μ;g/kg was obtained.
Conclusion
FI-ID-CV-ICP/MS has been successfully developed and applied to the determination of mercury in fly ash sample. The determined value of Hg in BCR 176R fly ash by the method of FI-CV-ID-ICP/MS described in this paper was well agreed with indicative result within the range of uncertainty. The difference of analytical results between digestion methods used two different acid mixtures was negligible. And also the difference between average signal processing mode and maximum mode was insignificant. The use of this method described in this paper is a considerable advantage in that it affords complete separation of mercury from the HF treated sample matrix and much higher sensitivity than conventional sample introduction method and reduced memory effect.
References
Mann J. L. , Long S. E. , Kelly W. R. 2003 J. Anal. At. Spectrom. 18 1293 -
Costley C. T. , Mossop K. F. , Dean J. R. , Gargen L. M. , Marshall J. , Carroll J. 2000 Anal. Chim. Acta 405 179 -
Armstrong H. L. , Corns W. T. , Stockwell P. B. , O’Connor G. , Ebdon L. , Evans E.H. 1999 Anal. Chim. Acta 390 245 -
Allibone J. , Fatemian E. , Walker P. J. 1999 J. Anal. Atom. Spectrom. 14 235 -
Milton M. J. T. , Quinn T. J. 2001 Metrologia 38 289 -
Jian L. , Goessler W. , Irgolic K. J. 2000 Fresenius’ J. Anal. Chem. 366 48 -
Karunasagar D. , Arunachalam J. , Gangadharan S. 1998 J. Anal. Atom. Spectrom. 13 679 -
Chen S. F. , Jiang S. J. 1998 J. Anal. Atom. Spectrom. 13 673 -
Liao H. C. , Jiang S. J. 1999 J. Anal. Atom. Spectrom. 14 1583 -
Long S. E. , Kelly W. R. 2002 Anal. Chem. 74 1477 -
Willie S. 2002 Am. Lab. 34 40 -
Mester M. , Angelone M. , Brunori C. , Cremisini C. , Muntau H. , Moraboto R. 1999 Analytica Chimica Acta 395 157 -
Woller Á. , Garraund H. , Martin A. , Donard O. F. X. , Fodor P. 1997 J. Anal. Atom. Spectrom. 12 53 -
De Bièvre P. 1994 J. Fresenius Anal. Chem. 350 277 -
1993 Guide to the expression of uncertainty in measurement: SBN 92-67-10188-9 1st ed. ISO Geneva, Switzerland
Tayor B. N. , Kuyatt C. E. 1994 Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results U. S. Government Printing Office Washington DC NIST technical note 1297
1998 EURACHEM Guide: The fitness for purpose of analytical methods. A laboratory guide to method validation and related topics 1st ed. Eurachem
Taylor H. E. 2001 Inductively coupled plasma mass spectrometry: Practice and techniques Academic Press