Determination of the Uranium Backgrounds in Lexan Films for Single Particle Analysis using FT-TIMS technique
Determination of the Uranium Backgrounds in Lexan Films for Single Particle Analysis using FT-TIMS technique
Mass Spectrometry Letters. 2011. Jun, 2(2): 57-60
  • Published : June 15, 2011
Download
PDF
Export by style
Article
Author
Metrics
Cited by
About the Authors
Park, Su-Jin
Park, Jong-Ho
Lee, Myung-Ho
Song, Kyu-Seok

Abstract
As background significantly affects measurement accuracy and a detection limit in determination of the trace amounts of uranium, it is necessary to determine the impurities in the Lexan detector film for single particle measurements by thermal ionization mass spectrometry coupled with fission track technique (FT-TIMS). We have prepared various micro sizes of the blank Lexan detector film using a micromanipulation technique for uranium measurements by TIMS. Few tens of fg of uranium background with no remarkable dependency on the film sizes were observed in the blank Lexan films with the sizes from $50{\times}50\;{\mu}m^2$ to $300{\times}300\;{\mu}m^2$. Based on the determination of the uranium background in the Lexan film, any background correction is necessary in the isotopic analysis of a uranium single particle with micron sizes when the particle bearing Lexan film is dissected with less than $300{\times}300\;{\mu}m^2$ size. The isotopic analysis of a uranium particle in U030 standard material using TIMS was carried out to verify the applicability of the Lexan film to the single particle analysis with high accuracy and precision.