Optical Method to Determine Gait Parameters Using Position Sensitive Detector
Optical Method to Determine Gait Parameters Using Position Sensitive Detector
Journal of Electrical Engineering and Technology. 2015. Sep, 10(5): 2155-2161
Copyright © 2015, The Korean Institute of Electrical Engineers
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
  • Received : October 10, 2013
  • Accepted : May 15, 2015
  • Published : September 01, 2015
Download
PDF
e-PUB
PubReader
PPT
Export by style
Article
Author
Metrics
Cited by
TagCloud
About the Authors
Gu-In Jung
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Ji-Sun Kim
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Tae-Hee Lee
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Ju-Hyeon Choi
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Han-Byeol Oh
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
A-Hee Kim
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Bong-Jun Goh
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Jun-Sik Kim
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Eun-Suk Lee
Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (guin83@konkuk.ac.kr, {kimji0330,t-a-e-h-e-e, wngus829, xhsl852, aheek, bey1212, nattee, eunsuk8097}@nate.com)
Jae-Hoon Jun
Corresponding Author : Dept. of Biomedical Engineering, BK21 Plus Research Institute of Biomedical Engineering, College of Biomedical & Health Science, Konkuk University, Chungju, Republic of Korea. (jjun81@kku.ac.kr)

Abstract
This study suggests an optical method to measure cardinal of gait (step width, step length, and stride length) with position sensitive detector (PSD). The effect of reflector’s shape (flat and cylinder) on the PSD output voltage was examined for the application of the suggested system to real situations with a curved shape reflector (e.g. shoes). Various mathematical models were evaluated to find the optimal equation for the distance measurement. Considering the effect of shape on detected signal, the inverse polynomial model was developed. The suggested method is simple to operate, low in cost, small in size, and can evaluate gait parameters in real time. This method is expected to be useful in the field of rehabilitation and sport science
Keywords
1. Introduction
For the rehabilitation process, measurements of the temporal and spatial features of gait are very important for monitoring functional recovery [1] . Quantitative gait analysis is required to evaluate the effects of various rehabilitation treatments designed for improving patient’s gait disability [2 - 5] .
Many methods have been studied to evaluate walking pattern. Measuring foot prints from patient’s painted feet was used to evaluate walking pattern. This method was simple but had limitations in space and accuracy. Walking kinematics can be evaluated with accelerometer but it can be shock-sensitive and has not enough accuracy due to double integration of acceleration for the derivation of displacement [6 - 10] .
The 3-D motion capture system with multiple IR cameras is widely used. However, the equipment is very expensive and can only be used in a specialized motion laboratory. To process data, specialists are required and the data processing time is relatively long. The system requires calibration in each test set and long preparation time. Sometimes patients claim inconvenience in attaching reflection makers on their body [11 - 12] .
Force plate method requires multiple force sensors to measure walking cycle within limited plate area [13] . This method has difficulty in evaluation walking pattern of patients with paralysis. Foot switch method can be used for detection of foot contacts on the ground. This method is useful for normal gait analysis but not suitable for pathological gait because of abnormal contact patterns such as loss of heel contact [14 - 17] .
The purpose of this study is to suggest alternative method for the identification of gait feature which would be convenient, inexpensive, and operating in real time. According to the principle of triangulation method, position sensitive detector (PSD) converts detect reflected light into distance-related voltage. After mathematical modeling of PSD output, we can obtain step width, step length, and stride length. Our final objective is to monitor gait parameters without time and space limitations using wireless module.
2. Materials and Methods
- 2.1 Gait features
Walking patterns can be described by gait phase, gait cycle, step length, and stride length. Stride length indicates the distance between two sequential initial contacts of the identical foot (e.g., right to right), step length indicates the distance between initial contact of one foot and the other foot (e.g., right to left), step width indicates the mediolateral distance between the locations of sequential initial contacts by two feet (see Fig. 1 ) [18] .
PPT Slide
Lager Image
Definition of gait parameters
- 2.2 Position sensitive detector (PSD)
Detection of distance in PSD is based on the principle of triangulation method [19 - 20] . PSD is not be affected by the color of reflector, while most optical sensors show change in the light intensity when reflected by different colors. PSD detects reflected light which induces electric current. The electric current ratio is determined by the hitting point of reflected light on the sensing area of light detector which converts the current ratio into voltage as an output signal.
Fig. 2 shows the principle of triangulation method in PSD sensor. Eq. (1) describes the relationship among lengths (L 1 , L 2 , X 1 , and X 2 ). The angle of incident beam can be changed by the ratios L 1 / L 2 and X 1 / X 2 . PSD sensor (GP2Y0A21YK0F, sharp, Japan) module is composed of light emitter (IR diode), lenses, and light detector. When the distance between sensor and reflectors is long, the incident beam hit the right part of the sensing area of light detector (see solid line). Likewise, when the distance between sensor and reflector is short, the incident beam hit the left part of the sensing area of light detector (see dashed line).
PPT Slide
Lager Image
PSD sensor’s signal detection principles
PPT Slide
Lager Image
where L 1 is the distance between lens 1 and reflecting object, L 2 is the distance between lens 2 and light detector (fixed), X 1 is the distance between lens1 and lens2 (fixed), X 2 is the distance between normal point to lens 2 and the hitting point of incident light in the sensing area of light detector.
The distance measuring range is 10 - 80 cm. It provides an analog output voltage that is correlated to the distance. Its supply voltage can be from 4.5 to 5.5 V and typical average dissipation current can be from 30 to 40mA.
- 2.3 The effect of reflector’s shape
For the application of PSD to real situations, curved shape reflector must be considered. The effect of reflector’s shape on PSD output voltage was investigated with specially designed setups. Reflecting objects such as ankle and shoes are not perfectly straight. They usually have curved shapes. Experimental setup was designed to test the effect of reflector’s shape (see Fig. 3 ). The diameter of cylindrical reflector was 8 cm which was similar to shoe’s heel shape. The PSD sensor is fixed on an optical table. The distance between the sensor and the reflectors was changed by an optical rail (PRL-24, Newport, Irvine, CA) with a resolution of 0.005 cm.
PPT Slide
Lager Image
Setup for reflector’s shape
The obtained results could be useful for realistic applications on shoes or ankles.
- 2.4 Data acquisition and calibration
Output signals from PSD sensor were collected using DAQ board (NI USB-6008, National Instrument, Austin, Texas) in digital form, and saved in a computer with LabVIEW (NI LabVIEW 8.0, National Instruments, Austin, Texas). Modeling of the experimental data was done by using Matlab (R2008a, Mathworks, Portola Valley, California).
Polynomial equations (1 st to 5 th order), exponential function, power function, logarithmic function, and fractional function were tested for calibration. Forward and inverse mathematical models were selected to obtain the distance. First, the relationship between PSD output and distance was mathematically modeled. Next, the inverse relationship was used to obtain the distance.
The inverse model is possible when the functions show monotonic pattern (increase or decrease). Eqs. (2)-(6) show inverse models with polynomial equations (1 st to 5 th order), exponential function, power function, logarithmic function, and fractional function.
PPT Slide
Lager Image
PPT Slide
Lager Image
PPT Slide
Lager Image
PPT Slide
Lager Image
PPT Slide
Lager Image
where a 0 ~ a n are coefficients of polynomial equation, b 0 ~ b 1 are coefficients of exponential function, c 0 ~ c 1 are coefficients of power function, d 0 ~ d 1 are coefficients of logarithmic function, e 0 ~ e 1 are coefficients of fractional function, x is the distance [cm], and y is the PSD output voltage [V].
- 2.5 Measurements of gait parameters
Fig. 4 shows schematic of the calculation of gait parameters. Two PSD sensors were used to measure the step width and the cross distance between toe and heel. The role of PSD 1 and PSD 2 were to obtain the cross distance between toe and heel (S 1 ) and step width (S 2 ), respectively. After, applying Eqns. (7)-(9), step width, step length, and stride length can be calculated and monitored with LabVIEW in real time
PPT Slide
Lager Image
Schematic for the calculation of gait parameters
Four healthy young adults (two females and two males, mean age = 25.75 ± 3.5 years old) participated in the study. The gait parameters such as step width, step length, and stride length were evaluated in static positions. To test the possibility of PSD sensor for gait analysis, human subjects wore the shoes attached with PSD sensor as shown in Fig. 4 .
PPT Slide
Lager Image
PPT Slide
Lager Image
PPT Slide
Lager Image
where S 1 is the cross distance between toe and heel [cm], S 2 is the step width [cm], S 3 is the normal distance between toe of one foot and heel of the other foot [cm], and fl is the foot length [cm].
To decrease the deviation of measuring distance by moving direction of the reflective object, we considered to set the sensor that the moving direction of the reflector and the line between emitter center and detector center are vertical. The sensors were mounted vertically and the shoes moved horizontally.
3. Results and Discussion
- 3.1 The effect of reflector’s shape
Fig. 5 shows the effect of reflector’s shape on PSD output. Flat and cylinder reflectors show some differences in short distance up to 15 cm, while differences are negligible in long distance. Various models were applied to find the optimal function to get distance.
PPT Slide
Lager Image
Effect of reflector’s shape
Table 1 shows coefficients of determination (R 2 ) and root-mean-square error (RMSE) with polynomial equations (1 st to 5 th order), exponential function, power function, logarithmic function, and fractional function. The results shows 5 th order polynomial equation has the optimal R 2 and the minimum RMSE for both flat and cylinder reflectors. Thus 5 th order polynomial function was used for inverse solution model to obtain distance.
Modeling results (R2and RMSE)
PPT Slide
Lager Image
Modeling results (R2 and RMSE)
Fig. 6 shows calibration results with inverse 5 th order polynomial equation for flat and cylinder reflectors. Only monotonically decreasing region was used to get inverse 5 th order equation. Inverse of 5 th order equation had coefficients of determination of flat (R 2 = 0.993 with p<.001) and cylinder reflectors (R 2 =0.9981 with p<.001).
PPT Slide
Lager Image
Inverse of 5th order polynomial
- 3.2 Measurements of gait parameters
The PSD sensors were attached on a pair of shoes and experiments were performed to find the optimal mathematical model for the use in gait analysis. In this experiment, PSD sensors were attached on one shoe and the other shoe was used as reflector.
Fig. 7 shows calibration results with inverse 5 th order polynomial equation of shoe model (R 2 = 0.991 with p<.001).
PPT Slide
Lager Image
Inverse of 5th order polynomial for shoe model
Fig. 8 shows relationship between the actual distance and LabVIEW output distance. The result showed actual distance was correlated well with calculated LabVIEW output distance. Experimental results showed that the RMSE was ±0.3185 cm.
PPT Slide
Lager Image
Actual distance versus calculated distance with shoe model
Fig. 9 shows LabVIEW display to get the distance from the inverse of 5 th order polynomial equation in real time. For measurements with human subjects wearing the sensor attached shoes, step length and stride length were calculated using gait parameters described in Eqns. (7)-(9) and Fig. 4 .
PPT Slide
Lager Image
Example of LabVIEW display
Fig. 10 shows the gait parameters such as step width, step length, and stride length in four subjects (two female and two male) in static positions. To test the possibility of PSD sensor for gait analysis, human subject wore the shoes attached with PSD sensor as shown in Fig. 4 . By measuring the cross distance between toe and heel (S 1 ) and step width (S 2 ), step length and stride length could be obtained using Eqns. (7)-(9). The relationship between the reference distance and the calculated distance output was evaluated. Experimental results with four subjects showed that the inverse 5 th order equation had RMSEs of gait parameters such as step width, step length, and stride length were ± 0.0135 cm, ± 0.7792 cm, and ± 1.5584 cm, respectively. Other models such as power function and fractional function (see Table 1 ) were also tested for possible practical use. Inverse power function had RMSEs of gait parameters such as step width, step length, and stride length were ± 0.1654 cm, ± 0.8419 cm, and ± 1.6839 cm, respectively. Also, inverse power function had RMSEs of gait parameters such as step width, step length, and stride length were ± 0.2436 cm, ± 1.0398 cm, and ± 2.0796 cm, respectively. Both power function and fractional function were good for possible use; however, the inverse 5 th order equation was the best model with the smallest RMSEs of gait parameters. Measured values were correlated with reference distances in test range.
PPT Slide
Lager Image
Measurement of gait parameters
The attached angles in the range of ± 4° deviation were tested and the RMSE of S 1 was ± 0.4449 cm. As results, the gait parameter errors such as the RMSEs of step width, step length, and stride length were ± 0.0135 cm, ± 0.7792 cm, and ± 1.5584 cm, respectively. The step width (S 2 ) is independent of S 1 while the step length and stride length are dependent on S 1 . The errors of the step length and stride length can be increased as the error of S 1 is increased.
Even though the optical method has limitation in accuracy, the suggested technique can be used conveniently using a remote module without space limitation and in real time using LabVIEW display. We suggested a new optical method to obtain gait parameters. In future study, we will reduce the S 1 error with multiple sensors and the better sensor design.
Our results are not perfectly accurate at this stage and have difficulty in collecting data during continuous dynamic motion. Wireless module should be added in future study for dynamic motion without limitations of testing space.
The suggested optical sensor is more durable than the accelerometer-based sensor. Even though the camera-based system has the high precision of human motion, the camera equipment is very expensive and can only be used in a motion laboratory with specialists performing data analysis. The proposed sensor can be used without space limitation and in real time with the LabVIEW program. Thus, the proposed sensor has advantages that other commercial products cannot offer.
At least, the results of this study imply that optical method using PSD sensor has high possibility to perform gait analysis. Further study and better design should be done to get data during dynamic motion. We may use multiple PSD sensors with ultrasound sensors for better accuracy. Wireless module to transfer the sensor output signal to PC should be done for monitoring gait parameters which may reduce time and space limitations. By differentiation of distance, further information about walking speed and acceleration may be obtained. The results can be applied in the field of rehabilitation and sport science.
4. Conclusion
In this study, we developed an optical method to measure step width, step length, and stride length using PSD sensors. This sensor is based on the principle of triangulation method. While most optical sensors show color dependency, PSD sensor is not affected by reflector’s color. Thus, we can collect distance data without color dependency and this is a very strong benefit as optical sensors.
The effect of reflector’s shape was investigated with flat and cylinder reflectors to get accurate distance information. Various mathematical models such as polynomial equations, exponential function, power function, logarithmic function, and fractional function were evaluated to find the optimal equation. With 5 th order polynomial equation, we obtained the best R 2 and the minimum RMSE. In this study, PSD sensors were attached on one shoe and the other shoe was used as reflector. The result showed that actual distance was correlated well with calculated LabVIEW output distance. The measured gait parameters with human subjects imply that our optical method can be used for gait analysis. This study can be applied in the field of rehabilitation and sport science.
Acknowledgements
This research was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education, Science and Technology (No. 2010-0023158) and the Pioneer Research Center Program through the National Research Foundation of Korea funded by the Ministry of Science, ICT & Future planning (No. 2011-0027920).
BIO
Gu-In Jung He received his B.S., M.S., and Ph.D. degrees in Biomedical Engineering from Konkuk University, Chungju, Korea, in 2008, 2010, and 2015, respectively. At present, he is an advanced researcher in the Laser Application Center at Kyungpook National University.
Ji-Sun Kim She received the B.S. degree and M.S. degree in Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2009 and 2011. She is currently working toward the Ph.D. degree in Biomedical Engineering with the college of Biomedical & Health Science, KonKuk University (KKU). Her research interests include laser tissue interaction and color detection.
Tae-Hee Lee She received the B.S. degree and M.S. degree in Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2011 and 2015. Her research interests include color distinguishing study and characterization of laser tissue interacttions.
Ju-Hyeon Choi She received the B.S. degree and M.S. degree in Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2013 and 2015. Her research interests include estimation of the light intensity by using LED sensory system, studies on interaction mechanism of laser.
Han-Byeol Oh He received the B.S. degree in Department of Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2014. He is currently working toward the M.S. degree in Department of Biomedical Engineering with the College of Biomedical and Health Science, KonKuk University (KKU). His research interests include simulation for optic, application of optical sensor, and studies on laser.
A-Hee Kim She received the B.S. degree in Department of Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2014. She is currently working toward the M.S. degree in Department of Biomedical Engineering with the College of Biomedical and Health Science, KonKuk University (KKU). Her research interests include angle measurement design using optical fiber, studies on laser.
Bong-Jun Goh He received the B.S. degree in Department of Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2015 He is currently working toward the M.S. degree in Department of Biomedical Engineering with the College of Biomedical and Health Science, KonKuk University (KKU). His research interests include application of optical sensor, and studies on laser.
Jun-Sik Kim He received the B.S. degree in Department of Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2015. He is currently working toward the M.S. degree in Department of Biomedical Engineering with the College of Biomedical and Health Science, KonKuk University (KKU). His research interests include angle measurement design using optical fiber, studies on laser.
Eun-Suk Lee She received the B.S. degree in Department of Biomedical Engineering from Konkuk University (KKU), Chungju, South Korea, in 2015. She is currently working toward the M.S. degree in Department of Biomedical Engineering with the College of Biomedical and Health Science, KonKuk University (KKU). Her research interests studies on laser and color detection.
Jae-Hoon Jun He received the B.S. degree in Chemical Engineering from Korea University, Seoul, KOREA, the M.S. degree in Chemical Engineering and the Ph.D. degree in Biomedical Engineering from Texas A&M University, College Station, TX, USA respectively. He was a Research Associate in Biomedical Engineering at Virginia Commonwealth University-Medical College of Virginia campus, Richmond, VA. He is currently a Professor in Biomedical Engineering at Konkuk University, Chungju, South Korea. His current research interests include optical sensing, photothermal therapy, tissue characterization, biomechanics, and T-ray imaging.
References
Sekin Masaki , Abe Yoshiyuki , Sekimoto Mitsuyoshi , Higashi Yuji , Fujimoto Toshiro , Tamura Toshiyo , Fukui Yasuhiro 2000 “Assessment of gait parameter in hemiplegic patients by accelerometry,” Proc. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 3 1879 - 1882
Lee Ju-Won , Lee Gun-Ki 2005 “Adaptive postural control for trans-femoral prostheses based on neural networks and EMG signals,” Int. J. Precis. Eng. Manuf. 6 (3) 37 - 44
Kim Gyoosuk , Kang Shinill , Kang Sungjae , Ryu Jeicheong , Mun Museoung , Kim Kyunghoon 2009 “Unlockable knee joint mechanism for powered gait orthosis,” Int. J. Precis. Eng. Manuf. 10 (3) 83 - 89    DOI : 10.1007/s12541-009-0051-y
Kong Se-Jin , Kim Chul-Seung , Moon Ki-Wook , Eom Gwang-Moon , Tack Gye-Rae , Kim Kyeong-Seop , Lee Jeong-Whan , Lee Young-Hee 2007 “Gait-Event Detection using an Accelerometer for the Paralyzed Patients,” Trans. KIEE 56 (5) 990 - 992
Moon Ki-Wook , Kim Chul-Seung , Lee Jae-Ho , Kwon Yu-Ri , Kang Dong-Won , Khang Gon , Kim Yo-Han , Eom Gwang-Moon 2009 “Development of Joint Angle Measurement System for the Feedback Control in FES Locomotion,” Trans. KIEE 58 (1) 203 - 209
Luinge Henk, J. , Veltink Peter H. 2004 “Inclination measurement of human movement using a 3-D accelerometer with autocalibration,” IEEE Trans. Neural Syst. Rehabil. Eng. 12 (1) 112 - 121    DOI : 10.1109/TNSRE.2003.822759
Mayagoitia Ruth E. , Nene Anand V. , Veltink Peter H. 2002 “Accelerometer and rate gyroscope measurement of kinematics: an inexpensive alternate to optical motion analysis systems,” J. Biom. 35 (4) 537 - 542    DOI : 10.1016/S0021-9290(01)00231-7
Morris J.R.W. 1973 “Accelerometry-A technique for the measurement of human body movement,” J. Biom. 6 (6) 729 - 736    DOI : 10.1016/0021-9290(73)90029-8
Marcuse Dietrich 1976 “Curvature lose formula for optical fibers,” J. Opt. Soc. Am. 66 (3) 216 - 220    DOI : 10.1364/JOSA.66.000216
Evans A.L. , Duncan G. , Gilchrist W. 1991 “Recording accelerations in body movements,” MED. Biol. Eng. Comput. 29 (1) 102 - 104    DOI : 10.1007/BF02446305
Bernard Auvinet , Denis Chaleil , Eric Barrey 1999 “Accelerometric gait analysis for use in hospital outpatients,” Rev. Rhum. Engl. Ed. 66 (7-9) 389 - 397
Whittle Michael W. 1982 “Calibration and performance of a 3-dimensional television system for kinematic analysis,” J. Biom. 15 (3) 185 - 196    DOI : 10.1016/0021-9290(82)90251-2
Park Sun Woo , Sohn Ryang Hee , Ryu Ki Hong , Kim Young Ho 2010 “Comparison of motion sensor system for gait phase detection,” J. Korean Sco. Precis. Eng. 27 (2) 145 - 152
Hausdorf Jeffrey M. , Ladin Zvi , We d Jeanne Y. 1995 “Footswitch system for measurement of the temporal parameters of gait,” J. Biom. 28 (3) 347 - 351    DOI : 10.1016/0021-9290(94)00074-E
Ann Seung Chan , Hwang Sung Jae , Kang Sung Jae , Kim Young Ho 2004 “Development and evaluation of a new gait phase detection system using FSR sensors and a gyrosensor,” J. Korean Sco. Precis. Eng. 21 (10) 196 - 203
Brandell Bruce R. 1982 “Development of a universal control unit for functional electrical stimulation (FES),” Am. J. Phys. Med. 61 (6) 279 - 301
Kong Se-Jun , Eom Gwang-Moon , Kim Chul-Seung , Park Kwan-Yong 2006 “Portable Gait-Event Detection System for FES Locomotion,” Trans. KIEE 55D (5) 248 - 253
Jacquelin Perry 1992 “Gait analysis, Normal and pathological function,” Slack Thorofare, NJ
Tong Q B , Ding Z L , Chen J C , Ai L L , Yuan F 2006 “The research of screw thread parameter measurement based on position sensitive detector and laser,” J. Phys.: Conf. Ser. 48 561 - 565    DOI : 10.1088/1742-6596/48/1/105
Jung Gu-In , Kim Ji-Sun , Hur Dong-Hun , Yu Hwan-Dong , Lim Sung-Hwan , Choi Ju-Hyeon , Lee Jeong-Hwan , Eom Gwang-Moon , Jun Jae-Hoon 2011 “Angular Displacement Measurement Using Optical Sensor,” Trans. KIEE 60 (10) 1959 - 1965