Simple Technique for Measurement of Complex Permittivity and Detection of Small Permittivity Change Using Partially Open Cavity
Simple Technique for Measurement of Complex Permittivity and Detection of Small Permittivity Change Using Partially Open Cavity
Journal of Electrical Engineering and Technology. 2014. Jan, 9(1): 268-272
Copyright © 2014, The Korean Institute of Electrical Engineers
  • Received : February 12, 2013
  • Accepted : August 09, 2013
  • Published : January 01, 2014
Download
PDF
e-PUB
PubReader
PPT
Export by style
Article
Author
Metrics
Cited by
TagCloud
About the Authors
Sangbok Park
Dept. of Electrical and Computer Engineering, University of Seoul, Korea. (sbpark10per@uos.ac.kr)
Young-seek Chung
Dept. of Radio Science and Engineering, Kwangwoon University, Korea. (yschung@kw.ac.kr)
Changyul Cheon
Corresponding Author: Dept. of Electrical and Computer Engineering, University of Seoul, Korea. (changyul@uos.ac.kr)

Abstract
This letter presents a measurement methodology of the complex permittivity of liquid using a partially open cavity in narrow band. The partially open cavity (POC) can measure dielectric small changes caused by the temperature variation of the liquid inside the cavity as well. Using the resonance frequency and unloaded quality factor of the proposed POC, the complex permittivity is evaluated. The apertures on the walls of the cavity are designed to circulate the liquid inside to outside of the POC and located at the corner area of the cavity to minimize the disturbance of field distribution at the dominant mode. The results measured by the proposed POC were compared with those by the conventional open-ended probe and Cole-Cole equation. The POC showed better performance in measuring small dielectric constant changes than the open-ended probe.
Keywords
1. Introduction
With several methods and probes of many types such as coaxial and waveguide transmission lines (T/L), and cavity, many people have researched to know electrical properties such as permittivity and permeability of materials at microwave frequencies [1 - 5] . If we can know electrical properties of the materials in an area of interest, they will be utilized to predict and analyze electromagnetic effects. And the small change of the permittivity of the materials in the area will affect the electric and magnetic fields distribution, which means that the small change should be measured to increase the accuracy of the field analysis. Most materials have the potential for the permittivity change as the degradation process occurs or the temperature of the materials changes at normal circumstance.
It is difficult for the conventional T/L probes to measure such small permittivity changes because the permittivity variations may be less than the tolerance of them. The technique using T/L probes for permittivity measurement cannot give precise results of the permittivity change of a material because the measurement error is generated during a calibration process using the reference materials such as air, methanol and distilled water. The measurement error is about 6% and depends on a measurement method and calibration skill [1] [2] . Also it would have the irreversible deformation of the probe filled with Teflon when measuring the material at over 40℃ [6] .
Thus, we may need a more sensitive probe of the cavity type having high quality factor. In order to detect the small permittivity change, a cavity resonator would be desirable since the resonance frequency of a cavity resonator with high quality factor is noticeably shifted even if the permittivity of the material inside the cavity gets changed even below 0.1%. Thus, we have proposed a partially open cavity (POC), which partially has openings to circulate liquid materials at each corner, to measure complex permittivity. If a vector network analyzer to measure the resonance frequency can implement a narrower span over measurement frequencies, one may detect the smaller permittivity changes [7] .
In this letter, the introduced partially open cavity probe can be applied to various monitoring systems for an insulating oil degradation, a temperature change of gas tank, and so on [8] [9]
2. Design of Partially Open Cavity
Fig. 1 shows the proposed rectangular POC fed by a coaxial line. It has four apertures at each corner wall whose opening area is 10 mm by t mm. We can put open apertures at each corner without disturbing the dominant mode field distribution of the rectangular cavity because the electric field intensities at each corner are weak enough. The apertures will allow the test material of liquid type to circulate easily inside to outside of the POC while measuring.
PPT Slide
Lager Image
Structure of partially open cavity
The return loss was simulated using HFSS according to the width (t) of the aperture, t=0~6 mm, and the resonant frequency of the dominant mode were compared in Fig. 2 . In the EM simulation using HFSS, the test material of the cavity was distilled water whose dielectric constant is 81. As shown in Fig. 2 , when t < 2mm, the resonant frequency of the dominant mode is almost same as that of t=0 which is the entirely closed cavity. As the aperture width t increases, the field distributions are disturbed by the apertures because the electromagnetic power is leaked thru the apertures. The wider aperture allows test materials of liquid type to circulate easily, but it causes us to get inaccurate permittivity of the materials due to the distorted field distribution of TE 101 mode and leakage power.
PPT Slide
Lager Image
Resonant frequencies of dominant mode according to t
Considering this trade-off, we have chosen the aperture width as 1 mm, which is about one twentieth of a guided wavelength of the 1 st resonance mode. We assume that the cavity with the width, 1 mm, is the same as the entirely closed cavity when considering the graph of Fig. 2 .
The picture of the proposed POC is shown in Fig. 3 . It was made of aluminum plate whose the thickness is 3 mm, and was designed based on the structure in Fig. 1 , but considered the thickness of aluminum plate.
PPT Slide
Lager Image
POC with t=1 mm; a=21 mm, b=16 mm, and d=26 mm.
3. Theory
- 3.1 Calculation for obtaining
Consider a rectangular cavity whose dimensions are a, b and d filled with dielectric material whose dielectric and magnetic constants are
PPT Slide
Lager Image
and
PPT Slide
Lager Image
, respectively. The resonant frequency of mnl mode of the cavity is given as (1).
PPT Slide
Lager Image
where k mnl is a resonant wave number [10] and m =1, n =0, and l =1. Eq. (1) can be expressed as (2) including a calibration factor, C f that is to calibrate the square root term including the physical length factors such as a, b, and d, by multiplying C f . It was difficult to precisely fabricate the POC, which is very small size, about 20 mm. Once we know the
PPT Slide
Lager Image
and
PPT Slide
Lager Image
of a reference liquid inside the rectangular cavity and the resonant frequency f are measured, the physical length error can be calibrated with the calibration factor, C f .
PPT Slide
Lager Image
The reference liquid used in this letter is distilled water whose complex permittivity is calculated by Cole-Cole Eq. [1] . Once the resonant frequency is measured with the rectangular cavity, the C f would be derived.
One of the main drawbacks of the proposed technique is that it can be applied in a narrow band because of the use of cavity resonance.
- 3.2 Calculation for obtaining loss tangent
In order to obtain loss tangent of a test material inside a rectangular cavity, we used the critical-points method [11] which is to obtain unloaded Q of a rectangular cavity including an electrical probe. Unloaded quality factor is that
PPT Slide
Lager Image
where Q is unloaded quality factor, Q c is quality factor with lossy conducting walls, and Q d is quality factor with lossy dielectric filling. The Q d is the inverse number of loss tangent (tan𝛿) and the Q c can be written as
PPT Slide
Lager Image
where
PPT Slide
Lager Image
The unloaded Q can be derived as
PPT Slide
Lager Image
If the denominator of the first term in (5) is large enough, the loss tangent, the second term, can be obtained by only deriving unloaded Q of a rectangular cavity.
4. Measurement
In this letter, we used saline as a test material that is a mixture of distilled water and NaCl and made three kinds of the test materials whose salinity are 0.3, 0.6 and 0.9%, respectively. The results,
PPT Slide
Lager Image
and tan𝛿, measured by the POC were compared with those by open-ended coaxial probe and Cole-Cole equation as shown in Table 1 .
Comparison of calculated and measured complex permittivity (at 23℃)
PPT Slide
Lager Image
Comparison of calculated and measured complex permittivity (at 23℃)
We immersed the POC and open-ended coaxial probe into test materials, saline, in a beaker, and measured the return loss of those probes in order to measure complex permittivity. The proposed POC has a calibration process using only one reference material, distilled water, as (2), whereas the open-ended coaxial probe used three materials, air, methanol and distilled water for the calibration. The measurement of complex permittivity using the proposed method showed the error of less than about 6% [1] .
And the POC can measure small change of
PPT Slide
Lager Image
with high accuracy as the temperature of a liquid in a beaker is rising, but the open-ended coaxial probe is difficult to identify the small change. We showed the fact in Fig. 4 . As mentioned, the open-ended coaxial probe is filled with a dielectric material of Teflon that can be easily deformed due to high temperature. Both of the POC and open-ended coaxial probe used distilled water of 20℃ for a calibration process and measured the distilled water of 20℃ to 30℃ and 25℃ to 60℃ as shown in Fig. 4(a) and (b) . Fig.4(a) shows the open-ended coaxial probe has the error in measuring dielectric constant even if the irreversible deformation of the dielectric material in the coaxial cable due to high temperature was not present. However the results measured from the proposed POC showed a good agreement with those calculated using Cole-Cole Equation in the temperature of 20℃ to 60℃.
PPT Slide
Lager Image
Measured and calculated ɛ r' as function of temperature
5. Conclusion
The measuring technique with a partially open cavity (POC) was introduced to measure complex permittivity in a narrow band and to monitor any small dielectric constant changes as the temperature is slowly rising; it also demonstrated its feasibility by comparing with the conventional technique using open-ended coaxial probe and Cole-Cole equation.
Acknowledgements
This research was supported by Basic Science Research Program through the National Research Foundation of Korea(NRF) funded by the Ministry of Education, Science and Technology (2009-0076300).
BIO
Sangbok Park He received the B.S. and M.S. degrees in school of electrical and computer engineering from the University of Seoul, Seoul, South Korea, in 2007 and 2009, respectively. He is currently working toward the Ph.D. degree at the University of Seoul, Seoul, South Korea. He is interested in microwave application for biological applications, bioelectromagnetics, and the design and analysis of microwave.
Young-seek Chung He received his B.S., M.S., and Ph.D. degrees in electrical engineering from Seoul National University, Seoul, Korea, in 1989, 1991, and 2000, respectively. From 1991 to 1996, he was with the Living System Laboratory, LG Electronics. From 1998 to 2000, he was a Teaching Assistant of electrical engineering with Seoul National University. From 2001 to 2002, he was with Syracuse University, Syracuse, NY. From 2003 to 2005, he has been a faculty member with the Department of Communication Engineering, Myongji University, Kyunggi, Korea. He is currently a Professor of the Department of Electronics Convergence Engineering, Kwangwoon University, Seoul, Korea. His current interests are umerical analysis and inverse scattering using time-domain techniques.
Changyul Cheon He was born in Seoul, Korea on April 5, 1960. He received the B.S. degree and the M.S. degree in electrical engineering from the Seoul National University, Seoul, Korea in 1983 and 1985, respectively, and the Ph.D. degree in electrical engineering from the University of Michigan at Ann Arbor in 1992. From 1992 to 1995, he was with the Department of Electrical Engineering, Kangwon National University, Chuncheon, Korea, as an Assistant Professor. He is currently an Professor of Electrical Engineering at the University of Seoul (UoS), Seoul, Korea. His group at UoS is currently involved with the design and analysis of microwave and millimeter-wave passive device using FEM, FDTD, and MoM techniques as well as analysis of scattering problem. He is also interested in microwave application for biological applications.
References
Nyshadham A. , Sibbald C.L. , Stuchly S.S. 1992 “Permittivity measurements using open-ended sensors and reference liquid calibration - An uncertainty analysis,” IEEE Trans. Microwave Theory & Tech. 40 (2) 305 - 314    DOI : 10.1109/22.120103
Park Sangbok , Chung Young-seek , Kwon Youngwoo 2009 “A novel two-port probe for complex permittivity and permeability measurement of liquid and powder type materials,” Microwave and Optical Technology Letters 51 (12) 2977 - 2979    DOI : 10.1002/mop.24793
Al-Moayed Nawaf N. , Afsar Mohammed N. , Khan Usman A. , McCooey Sean , Obol Mahmut 2008 “Nano Ferrites Microwave Complex Permeablility and Permittivity Measurements by T/R Technique in Waveguide,” IEEE Trans. Magnetics 44 (7) 1768 - 1772    DOI : 10.1109/TMAG.2008.920846
Akhtar M. Jaleel , Feher Lambert E. , Thumm Manfred 2009 “Noninvasive procedure for measuring the complex permittivity of resins, catalysts, and other liquids using a partially filled rectangular waveguide structure,” IEEE Trans. Microwave Theory & Tech. 57 (2) 458 - 470    DOI : 10.1109/TMTT.2008.2011177
Nishikata Atsuhiro 2007 “A swept-frequency measurement of complex permittivity and complex permeability of a columnar specimen inserted in a rectangular waveguide,” IEEE Trans. Microwave Theory & Tech. 55 (7) 1554 - 1567    DOI : 10.1109/TMTT.2007.900340
Kim Namgon , Yoon Jeonghoon , Cho Sungjoon , Cho Jeiwon , Cheon Changul , Kwon Youngwoo 2008 “An optimum design methodology for planar-type coaxial probes applicable to broad temperature permittivity measurements,” IEEE Trans. Microwave Theory & Tech. 56 (3) 684 - 692    DOI : 10.1109/TMTT.2008.916986
Chua Lye Heng , Mirshekar-Syahkal Dariush 2003 “Accurate and direct characterization of high-Q microwave resonators using one-port measurement,” IEEE Trans. Microwave Theory & Tech. 51 (3) 978 - 985    DOI : 10.1109/TMTT.2003.808711
Gradnik Maja Koncan 2002 “Physical-chemical oil tests, monitoring and diagnostics of oil-filled transformers,” 2002 International Conference on Dielectric Liquids 242 - 247
Park Sangbok , Chung Young-seek , Cheon Changyul 2010 “A new method for measuring degradation level of insulating oil with temperature in microwave,” IEEE 2010 Antennas and Propagation Society International Symposium 1 - 4
Pozar D.M. 2003 Microwave engineering 3rd ed. J. Wiley & Sons New York
Sun En-Yuan , Chao Shuh-Han 1995 “Unloaded Q measurement - the critical-points method,” IEEE Trans. Microwave Theory & Tech. 43 (8) 1983 - 1986    DOI : 10.1109/22.402290