Development Rapid Analytical Methods for Inositol as a Trace Component by HPLC and LC-MS/MS in Infant Formula
Development Rapid Analytical Methods for Inositol as a Trace Component by HPLC and LC-MS/MS in Infant Formula
Food Science of Animal Resources. 2015. Aug, 35(4): 466-472
Copyright © 2015, Korean Society for Food Science of Animal Resources
This is an open access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licences/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
  • Received : January 19, 2015
  • Accepted : May 14, 2015
  • Published : August 31, 2015
Download
PDF
e-PUB
PubReader
PPT
Export by style
Article
Author
Metrics
Cited by
TagCloud
About the Authors
Jin-Ho, Shin
Jung-Min, Park
Ha-Jung, Kim
Jang-Hyuk, Ahn
KOTITI Testing & Research Institute, Seongnam 462-807, Korea
Byung-Man, Kwak
Food Safety Center, Research and Development Institute, Namyang Dairy Co., Ltd., Gongju 339-914, Korea
Jin-Man, Kim
jinmkim@konkuk.ac.kr

Abstract
A rapid and simple analytical method, using liquid chromatography tandem mass spectrometry (LC-MS/MS), was developed to detect myo -inositol (MI) in infant formulas. For protein removal: acid hydrolysis and lipid removal through organic solvent extraction. The operating conditions for instrumental analysis were determined based on previously reported analogous methods that used LC-MS/MS. Quantitative analysis was used for the detection limit test, infant formula recovery test, and standard reference material (SRM) 1849a to verify the validity of our LC-MS/MS analytical method, which was developed to quantify MI. For validation, the results of our method were compared with the results of quantitative analyses of certified values. The test results showed that the limit of detection was 0.05 mg/L, the limit of quantitation was 0.17 mg/L, and the method detection limit was 17 mg/kg. The recovery test exhibited a recovery between 98.07-98.43% and a relative standard deviation between 1.93-2.74%. Therefore, the result values were good. Additionally, SRM 1849a was measured to have an MI content of 401.84 mg/kg and recovery of 98.25%, which is comparable to the median certified value of 409 mg/kg. From the aforementioned results, we judged that the instrumental analysis conditions and preparation method used in this study were valid. The rapid analytical method developed herein could be implemented in many laboratories that seek to save time and labor.
Keywords
Introduction
The isomer myo -inositol (MI) is a meso compound with an optically inactive plane of symmetry; meso -inositol is an obsolete name that refers to MI. In addition to MI, the other naturally occurring stereoisomers (in minimal quantities) are scyllo- , muco- , D- chiro- , and neo -inositol; other possible isomers are L- chiro- , allo- , epi- , and cis -inositol. D- and L- chiro -inositol constitute the only pair of inositol enantiomers; they are enantiomers of each other, not of MI.
Inositol and some of its mono- and polyphosphate derivatives function as the basis for various signaling and secondary messenger molecules. They are involved in the following biological processes: insulin signal transduction ( Larner, 2002 ), cytoskeleton assembly, nerve guidance (epsin), intracellular calcium (Ca 2+ ) concentration control ( Gerasimenko ., 2006 ), cell membrane potential maintenance ( Kukuljan , 1997 ), breakdown of fats ( Rapiejko .,1986 ), and gene expression ( Shen ., 2003 ; Steger ., 2003 ). MI is a compound necessary for infant development and when this substance cannot be administered through breast milk, it must be supplied in the diet by either inositol-enriched milk or formula ( Koletzko ., 2013 ). Maximum limits permitted for MI in infant formulas range from 4.0 to 40.0 mg/100 kcal (1.0-9.5 mg/100 kJ), while the requirements for premature infants range from 27.0 to 67.5 mg/100 kcal ( Codex Alimentarius, FAO WHO, 2007 ). Effective quantification of MI is necessary to ensure that its amounts in powdered milk or infant formula do not exceed the established limits. At present, since there are numerous MI-enriched food products, rapid and simple methodologies to determine MI content are desirable. Various analytical methods have already been published to determine quantities of inositol: microbiological assays ( Difco Manual, 1984 ), gas chromatography ( Troyano ., 1996 ), high performance liquid chromatography ( Hicks ., 1994 ; Kargacin ., 1987 ; Lauro ., 1989 ), and ion chromatography ( Cataldi ., 1998 ). However, the preparation of samples containing emulsified foods, such as infant formulas, is challenging. Generally, the preparation method includes either alkali decomposition or fat-free processes; these were not considered in a previous report of inositol analysis. This study aims to introduce a sample preparation method that takes in consideration the matrix characteristics of formulated milk powder. This study also discusses the development of a simple test method that requires a smaller amount of test materials and time than the conventional method. The method presented herein is considered an improvement upon the AOAC journal method due to its precise instrumental analysis, simple sample preparation method, and utility of LC-MS/MS.
Materials and Methods
- Standard, Samples, and reagents
MI (Cat. No. 1340960) was purchased from the US Pharmacopeial Convention (USP, USA) for use as the reference standard material. The purity of MI was 99.9% (0.999 mg/mg). A stock solution of 1,000 µg/mL inositol was dissolved in water and diluted to concentrations of 1, 5, 10, 20, and 50 ng/mL to make the standard working solutions. The infant and toddler formulas used in this study were purchased from a local market and stored at 4℃. A certified reference material (CRM), infant formula SRM 1849a (NIST, USA), was used in the recovery tests to develop the method. The amount of inositol in SRM 1849a was 409 mg/kg. Both of 0.1 M hydrochloric acid (HCl) and ammonium acetate were purchased from Junsei Chemical (Japan). Water, methanol, and chloroform (HPLC grade) were purchased from Merck (Germany). Ultrapure water was obtained using a Banstead Diamond TII system (USA). The distilled water had a resistance of 18.0 MΩ.
- Sample preparation
A sample preparation method based on the previous study ( Schimpf ., 2012 ) was modified to include the following techniques for protein removal: acid hydrolysis and lipid removal through organic solvent extraction. A sample containing 1 g of infant formula was placed in a 100 mL volumetric flask and dissolved in 15 mL of distilled water and 1 mL of 0.1 M HCl. The mixture was agitated for 5 min. Afterwards, 10 mL of chloroform was added to remove the lipid fraction ( Sullivan and Carpenter, 1993 ) and the solution was covered with a screw cap and vigorously mixed for 1 min in a vortex mixer at maximum speed. The tube was then centrifuged for 10 min at 4,000 rpm at 4℃. The aqueous layer was filtered through a 0.2 µm nylon syringe filter and analyzed by LC-ESI-MS/MS.
- Operating conditions
The operating conditions for instrumental analysis were determined based on previously reported analogous methods that used LC-MS/MS ( Kim ., 2012 ). An Agilent 1200 HPLC system (Agilent, USA) equipped with a Prevail Carbohydrate ES column (4.6 mm × 250 mm, 5 µm, Alltech) and a 6410 triple quadrupole LC/MS tandem MS system were used to analyze inositol. The mobile phase was composed of 5 mM ammonium acetate (25%) and acetonitrile (75%) for isocratic flow. The following parameters were employed: 1.0 mL/min flow rate, 30℃ column temperature, and 10 µL injection volume. HPLC grade solvents were filtered through a 0.45 µm membrane and ultrasonically degassed prior to use. The specific conditions used in LC-MS/MS analysis are shown in Table 1 .
Liquid chromatography-tandem mass spectrometry conditions formyo-inositol
PPT Slide
Lager Image
Liquid chromatography-tandem mass spectrometry conditions for myo-inositol
- Validation of the method
The developed method was validated by the AOAC guidelines for single laboratory validation of chemical methods ( AOAC International, 2002 ). The method validation evaluated the following parameters: linearity, limit of detection (LOD), limit of quantification (LOQ), method detection limit (MDL), and recovery test results. Linearity was characterized by the average coefficient of determination (r 2 ) and was calculated using five consecutive standard curves. The LOD and LOQ were determined by diluting standard working solutions of inositol to obtain signal to noise ratios of ~3:1 for LOD and ~10:1 for LOQ. MDL was determined by multiplying the solvent volume (mL) of the LOD and dividing by the sample amount (g). The recovery tests of inositol were conducted with a spiking test involving an infant formula sample and confirmation of the certificated value for standard reference material (SRM) 1849a. For the spiking test, the spiked level of infant formula was 50 mg/kg for 0.3 g of sample and the results from quantitative analysis were compared to those obtained for samples without the standard solution
- Monitoring test
Infant formulas consisting of milk-based powder and cereal-based powder were analyzed for MI content. A total of 16 samples were used for the monitoring tests. All samples were purchased from a local market and stored at room temperature.
Results and Discussion
- Establishing the optimum conditions of the ESI ion source
According to a study published in the journal, the polarity of the existing ESI mass detector was set when the inositol assay was analyzed in the positive ( Flores ., 2011 ) or negative mode ( Kim ., 2012 ). In this study, the test was carried out in both positive and negative modes to determine a suitable polarity for the ESI mass detector used in the inositol LC-MS/MS assays.
It is worth noting that the precursor ion [M + H] (m/z = 181.2) of myo -inositol was not detected in positive mode. Meanwhile, the precursor ion [M-H] (m/z = 179.2) was detected in negative mode. Therefore, negative mode was used in the mass spectrometry process ( Fig. 1 ).
PPT Slide
Lager Image
Test results for optimal ESI mass detector’s polarity mode as a myo-inositol trace component by LC-MS/MS.
- Optimum condition settings to detect precursor and product ions
In negative mode, the precursor ion (179.2 m/z) was search for the quantitative and qualitative condition among many product ions. The results are shown in Fig. 2 . Showed 86.9 m/z is the highest sensitivity in Product ion, the second is 70.9 m/z, the third was a result of the high sensitivity to 98.8 m/z. In this study, we selected m/z = 86.9 as a quantitative ion, of the ion dose and the highest sensitivity to select the 98.8 m/z as qualitative ion. The 70.9 m/z ion is higher than 98.8 m/z from the surface sensitivity, sample analysis TIC (total ion chromatogram) and EIC (extract ion chromatogram) were detected in the peak that interference, which results in a high concentration of the sample during analysis was determined to be affected.
PPT Slide
Lager Image
Test results for quantitative and qualitative ion conditions of the product ion for ion precursor ion.
- Linearity and range
The MI content in infant formula was quantified to be approximately 409 mg/kg, and the certified value range of SRM 1849a was 401.84 ± 0.74 mg/kg. In case of completing sample preparation with final volume of 10 mL in comparison with sample of 0.3 g proposed by this study, the calibration curve was predicted to be within the range of 3-5 mg/L (ppm). The test results are shown in Fig. 3 and include an r 2 value of 0.9993; linearity was observed in solutions at concentrations of 1, 5, 10, 20, and 50 µg/L. Therefore, we employed solutions at concentrations within the range of 1-50 µg/L for the calibration curves in this test, and accordingly diluted the final solutions 200-fold for instrumental analysis.
PPT Slide
Lager Image
Monitoring test for myo-inositol in certified reference material (SRM 1849a) and infant formula.
- Method validations
Quantitative analysis was used for the detection limit test, infant formula recovery test, and certified reference material SRM 1849a to verify the validity of our LC-MS/MS analytical method, which was developed to quantify MI. For validation, the results of our method were compared with the results of quantitative analyses of certified values. The test results showed that the limit of detection (LOD) was 0.05 mg/L, the limit of quantitation (LOQ) was 0.17 mg/L, and the method detection limit (MDL) was 17 mg/kg. The recovery test exhibited a recovery between 98.07-98.43% and a relative standard deviation between 1.93-2.74%. Therefore, the result values were good. Additionally, SRM 1849a was measured to have an L-carnitine content of 401.84 mg/kg and recovery of 98.25%, which is comparable to the median certified value of 409 mg/kg. From the aforementioned results, we judged that the instrumental analysis conditions and preparation method used in this study were valid. The specific validation factors are shown in Table 2 .
Validation factors and monitoring test formyo-inositol in certified reference material (SRM 1849a) and infant formula using liquid chromatography-tandem mass spectrometry analysis
PPT Slide
Lager Image
Validation factors and monitoring test for myo-inositol in certified reference material (SRM 1849a) and infant formula using liquid chromatography-tandem mass spectrometry analysis
- Monitoring test for infant and toddler formulas
A monitoring test was carried out for 16 samples of infant formula and toddler formula with SRM 1849a, international certified reference material. According to the results of the monitoring test, which are shown in Table 2 , the MI content in powder products was quantified at levels between 134.85-1118.08 mg/kg. These levels were accurate from comparison with the contents indicated on the products. Therefore, it was possible to verify that MI indication management of infant formula placed on the Korean market was well accomplished.
Quantity corresponding to appropriate sample volume of each food group was 0.3 g for all sample groups and it was quantity corresponding to sample volume included in calibration curve concentration established by the method. The LC-MS/MS chromatogram and mass spectrum for MI are shown in Fig. 4 . The chromatogram and the mass spectrum showed that the standard solution, infant formula and SRM 1849a sample treated by the developed test method and standard solution.
PPT Slide
Lager Image
The LC-MS/MS chromatogram and mass spectrum for MI.
Conclusions
A method employing LC-MS/MS and a simple sample pretreatment procedure was developed to accurately quantify MI in infant formulas. We carried out this study along to suggested material and method. Sample pretreatment time and labor were reduced, and the recovery test exhibited accurate results for infant formula samples. The precursor ion for MI was detected at m/z = 179.2, and product ions were detected at m/z = 86.9 (quantitative) and m/z = 98.8 (qualitative), respectively. The results from the spiked recovery test were in the range of 93.18-95.31% and exhibited relative standard deviations between 1.93% and 2.74%; additionally, the result for certified reference material (SRM 1849a) was within the range of the certificated values. The developed method based on LC-MS/MS in MRM mode, following the described sample preparation, could be an accurate tool to replace conventional analytical methods, as well as reduce time and labor. Moreover, a beginner can be expected to easily perform this analytical procedure due to its simplicity.
Acknowledgements
This paper was supported by Konkuk University in 2014.
References
2002 AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals Association of Official Analytical Chemists Gaithersburg, MD, USA
Cataldi T. R. I. , Margiott G. , Zambonin C. G. 1998 Determination of sugars and alditols in food samples by HPAEC with integrated pulsed amperometric detection using alkaline eluents containing barium or strontium ions Food Chem. 62 109 - 115    DOI : 10.1016/S0308-8146(97)00154-4
Codex guideline, Draft Guidelines for the Design and Implementation of National Regulatory Food Safety Assurance Programs Associated with the Use of Veterinary Drugs in Food Producing Animals Codex Alimentarius Commission Rome, Italy
1984 Difco Manual, Difco Laboratories 10th ed. Detroit, MI 1093 -
Flores M. I. A. , Moreno J. L. F. , Frenich A. G. , Vidal J. L. M. 2011 Fast determination of myo-inositol in milk powder by ultra performance liquid chromatography coupled to tandem mass spectrometry Food Chem. 129 1281 - 1286    DOI : 10.1016/j.foodchem.2011.05.045
Gerasimenko J. V. , Flowerdew S. E. , Voronina S. G. , Sukhomlin T. K. , Tepikin A. V. , Petersen O. H. , Gerasimenko O. V. 2006 Bile acids induce Ca2+ release from both the endoplasmic reticulum and acidic intracellular calcium stores through activation of inositol trisphosphate receptors and ryanodine receptors J. Biol. Chem. 281 40154 - 40163    DOI : 10.1074/jbc.M606402200
Hicks K. B. , Hotchkiss A. T. , Sasaki K. , Irwin P. L. , Doner L. W. , Nagahashi G. , Haines R. M. 1994 Analytical and preparative HPLC of carbohydrates: Inositols and oligosaccharides derived from cellulose and pectin Carbohydr. Polym. 25 305 - 313    DOI : 10.1016/0144-8617(94)90056-6
Kargacin M. E. , Bassell G. J. , Ryan P. J. , Honeyman T. W. 1987 Separation and analysis of fluorescent derivatives of myo-inositol and myo-inositol 2-phosphate by high-performance liquid chromatography J. Chromatogr. A. 393 454 - 458    DOI : 10.1016/S0021-9673(01)94244-X
Kim B. H. , Park J. Y. , Jang J. B. , Moon D. C. 2012 LC-MS/MS method for the quantification of myo- and chiroinositol as the urinary biomarkers of insulin resistance in human urine Biomed. Chromatogr. 26 429 - 433    DOI : 10.1002/bmc.1682
Koletzko B. , Bhutta Z. A. , Cai W. , Cruchet S. , El Guindi M. , Fuchs G. J. , Goddard E. A. , van Goudoever J. B. , Quak S. H. , Kulkarni B. , Makrides M. , Ribeiro H. , Walker A. 2013 Compositional requirements of follow-up formula for use in infancy: Recommendations of an international expert group coordinated by the early nutrition academy Ann. Nutr. Metab. 62 44 - 54
Kukuljan M. , Vergara L. , Stojilkovic S. S. 1997 Modulation of the kinetics of inositol 1,4,5-trisphosphate-induced [Ca2+] ioscillations by calcium entry in pituitary gonadotrophs Biophys. J. 72 698 - 707    DOI : 10.1016/S0006-3495(97)78706-X
Larner J. 2002 D-chiro-inositol--its functional role in insulin action and its deficit in insulin resistance Int. J. Exp. Diabetes Res. 3 47 - 60    DOI : 10.1080/15604280212528
Lauro P. N. , Craven P. A. , DeRubertis F. R. 1989 Two-step high-performance liquid chromatography method for the determination of myo-inositol and sorbitol Anal. Biochem. 178 331 - 335    DOI : 10.1016/0003-2697(89)90648-9
Rapiejko P. J. , Northup J. K. , Evans T. , Brown J. E. , Malbon C. C. 1986 G-proteins of fat-cells. Role in hormonal regulation of intracellular inositol 1,4,5-trisphosphate Biochem. J. 240 35 - 40    DOI : 10.1042/bj2400035
Schimpf K. , Thompson L. , Baugh S. 2012 Determination of myo-inositol (free and bound as phosphatidylinositol) in infant formula and adult nutritionals by liquid chromatography/pulsed amperometry with column switching: First action 2011.18 J. AOAC Int. 95 937 - 942    DOI : 10.5740/jaoacint.CS2011_18
Shen X. , Xiao H. , Ranallo R. , Wu W. H. , Wu C. 2003 Modulation of ATP-dependent chromatin-remodeling complexes by inositol polyphosphates Science 299 112 - 114    DOI : 10.1126/science.1078068
Steger D. J. , Haswell E. S. , Miller A. L. , Wente S. R. , O'Shea E. K. 2003 Regulation of chromatin remodeling by inositol polyphosphates Science 299 114 - 116    DOI : 10.1126/science.1078062
Sullivan D. M. , Carpenter D. E. 1993 Methods of analysis for nutrition Labeling AOAC International Arlington, VA, USA Chapter 5 Lipid Analysis 85 - 103
Troyano E. , Villamiel M. , Olano A. , Sanz J. , Martinez-Castro I. 1996 Monosaccharides and myo-inositol in commercial milks J. Agr. Food Chem. 44 815 - 817    DOI : 10.1021/jf950260d